RT Book, Section T1 Spectrometric characterization of amorphous silicon PIN detectors A1 Leyva, A. A1 Ramirez, F. J. A1 Ortega Villafuerte, Yanicet A1 Estrada, M. A1 Cabal, A. A1 Cerdeira, A. A1 Díaz, A AB During the last years, much interest has been dedicated to the use of amorphous silicon PIN diodes as particle and radiation detectors for medical applications. This work presents the spectrometric characterization of PECVD high deposition rate diodes fabricated at our laboratory, with thickness up to 17.5 μm. Results show that the studied devices detect the Am^(241) alpha particles and the medical X-rays generated by a mammograph model Senographe 700T from General Electric. Possible reasons of the observed energy losses are discussed in the lest. Using the SRIM2000 program, the transit of 5.5 MeV alpha particles through a diode was simulated, determining the optimum thickness for these particles to deposit their energy in the intrinsic layer of the diode. PB Amer Inst Physics SN 1-56396-963-7 YR 2000 FD 2000 LK https://hdl.handle.net/20.500.14352/61005 UL https://hdl.handle.net/20.500.14352/61005 LA eng NO © Amer Inst Physics.Mexican Symposium on Medical Physics (4. 2000. Merida, Mejico).Authors would like to thank the Ionizing Radiation Metrology Laboratory of ININ for the use of his standards, X-ray units and installations. This work was supported by CONACYT project 28092A. NO CONACYT DS Docta Complutense RD 27 abr 2024