RT Journal Article T1 Evaluation of repeatability and agreement of two optical biometers for intraocular lens power calculation A1 Martínez Alberquilla, Irene A1 Svensson, Saga A1 Ruiz Alcocer, Javier A1 Madrid Costa, David A1 Dominguez-Vicent, Alberto A1 Priya Venkataraman, Abinaya AB The repeatability of two biometers (Lenstar-LS900 and Eyestar-900) to measure ocular parameters and intraocular lens (IOL) power calculation, and their agreement were evaluated. 134 eyes of 134 participants were measured thrice with each biometer. Axial length (AL), anterior chamber depth (ACD), lens thickness (LT) and keratometry (K) were evaluated. The IOL power was calculated using different formulas. The repeatability limit (RLimit), the mean differences (MD) and the limits of agreement (LoA) were calculated. The RLimits for all parameters were higher with Lenstar compared to Eyestar. RLimits were lower than 0.50 D except for Barrett Universal II (0.54 D) and Haigis (0.51 D) formulas with the Lenstar. Mean differences were lower than 0.01 mm for AL, ACD and LT, and lower than 0.03 D for K. MD ranged from 0 to 0.02 D for all formulas except for Barrett and Hill. When dividing the sample into subgroups (short, normal and long eyes), the MDs were similar for the IOL power and were lower than 0.03 D, except for the Barrett and Hill formulas. Both biometers provide repeatable biometry and IOL power calculations. The LoA interval for the IOL power calculation was between 0.75 and 1.50D, which was similar among the subgroups. YR 2024 FD 2024-09-27 LK https://hdl.handle.net/20.500.14352/109699 UL https://hdl.handle.net/20.500.14352/109699 LA eng NO Alberquilla, I.M., Svensson, S., Ruiz-Alcocer, J. et al. Evaluation of repeatability and agreement of two optical biometers for intraocular lens power calculation. Sci Rep 14, 22151 (2024). https://doi.org/10.1038/s41598-024-73206-0 NO Karolinska Institute DS Docta Complutense RD 10 abr 2025