RT Journal Article T1 Dosimetric study of bevel factors in IOERT with mobile linacs: towards a unified code of practice A1 Ayala, Rafael A1 García, Rocío A1 Ruiz, Gema A1 García, María Jesús A1 Soza, Álvaro A1 Gómez, Susana A1 Udías Moinelo, José Manuel A1 Ibáñez García, Paula Beatriz AB Background: Dosimetry in intraoperative electron radiotherapy (IOERT) poses distinct challenges, especially with inclined applicators deviating from international protocols. Ion recombination in ionization chambers, electron beam degradation due to scattering in cylindrical applicators, coupled with a lack of a welldefined beam quality surrogate, complicate output factor determination with ionization chambers. Synthetic diamond-based detectors, offer potential solutions; however, their suitability requires further exploration. Purpose: This study addresses output factor determination for beveled applicators. Objectives include assessing the suitability of PTW microDiamond detectors and determining correction factors for ionization chamber measurements based on energy variations at the depth of maximum dose ( z max ) for beveled applicators. Experimental data are compared against results obtained from Monte Carlo simulations. Methods: We conducted measurements using both PTW microDiamond and IBA CC01 detectors. In addition to benchmarking bevel factors with penEasy, we employed Monte Carlo simulations to determine angular response correction factors for microDiamond detectors and to evaluate energy variations at z max for beveled applicators. Results: The findings indicate that angular response correction factors are unnecessary for microDiamond detectors with beveled applicators. However, variations in mean energy at z max potentially impact absorbed dose calculations with ionization chambers, particularly with the most inclined applicators. Conclusions: Based on our results, the study recommends using microDiamond detectors over cylindrical ionization chambers for output factor determination in IOERT with inclined applicators. Addressing energy variations at z max is crucial to improve accuracy in ionization chamber measurements. These findings have implications for dosimetry protocols in IOERT, contributing to enhanced delivery in clinical practice. PB Elsevier SN 1120-1797 YR 2024 FD 2024-11 LK https://hdl.handle.net/20.500.14352/134947 UL https://hdl.handle.net/20.500.14352/134947 LA eng NO Ayala R, García R, Ruiz G, García MJ, Soza Á, Gómez S, et al. Dosimetric study of bevel factors in IOERT with mobile linacs: Towards a unified code of practice. Physica Medica 2024;127:104836. https://doi.org/10.1016/j.ejmp.2024.104836. NO © 2024 Associazione Italiana di Fisica Medica e Sanitaria. DS Docta Complutense RD 28 abr 2026