RT Journal Article T1 Approximate methods to determine the modulation transfer function of a hololens system: a comparative study A1 Varela Vargas, Alberto J. A1 Calvo Padilla, María Luisa AB We present a comparative study between two experimental methods to determine the modulation transfer function (MTF) of a hololens system. The two hololenses were previously recorded and tested for filtering pseudocolor. In the first method we used the classical Foucault test. The second, alternative method is based on the digital image processing of a perfect edge under incoherent illumination. From the digitized intensity line profiles we obtain the MTF and cutoff frequency of the optical system according to the reciprocity between line spread function and MTF. Comments are made on the applicability and accuracy of these two methods. PB Spie-Soc Photo-Optical Instrumentation Engineers SN 0091-3286 YR 1995 FD 1995-04 LK https://hdl.handle.net/20.500.14352/59082 UL https://hdl.handle.net/20.500.14352/59082 NO © 2005 COPYRIGHT SPIE--The International Society for Optical Engineering. DS Docta Complutense RD 11 abr 2025