RT Journal Article T1 Comparison of two methods to measure permeability of dentin A1 Macorra García, José Carlos De La A1 Escribano, Nuria Isabel AB Abstract: Dentin permeability was measured alternatively with two methods: a 10-ul capillary method with visual evaluation (PC) and a motorized automatic measuring device (Flodec, FD), both interposed in a simulated perfusion system. Eight human third molar coronal fragments were connected to systems, and their permeability to distilled water measured at 0, 5, 10, 15, 20, 25, and 29 cmH2O pressure. Resultant permeabilities (in ul/s) for both techniques were interrelated with the use of the Passing and Bablok nonparametric method, which gives information about the range of constant and proportional errors and their 95% confidence intervals (95CI). The relationship between the methods is described by the regression formula: FD =-0.0003 + 0.945·PC, with 95CI for constant (-0.0015–0.0009) and for slope (0.738–1.168), indicating that both methods are interchangeable, although not identical. YR 2002 FD 2002 LK https://hdl.handle.net/20.500.14352/60352 UL https://hdl.handle.net/20.500.14352/60352 LA spa DS Docta Complutense RD 10 abr 2025