RT Journal Article T1 Intrasession repeatability and agreement of the anterior corneal assessment provided by a multidiagnostic device A1 Barrio De Santos, Ana Rosa A1 Antona Peñalba, Beatriz A1 Morago Villanueva, Sofía A1 Martínez Arribas, Víctor A1 González Pérez, Mariano AB Clinical relevance: Multidiagnostic systems have recently appeared on the market. Knowledge of the repeatability and validity of any instrument is mandatory before its introduction in clinical practice.Background: The aim of this work is to examine the intrasession repeatability of anterior pole measurements provided by the multidiagnostic device Wave Analyzer Medica 700 (WAM700) and agreement with Pentacam measurements in normal eyes.Methods: In the right eyes of 113 participants, three repeat measurements of central keratometry, central corneal thickness, anterior chamber depth and corneal eccentricity were made with the WAM700 and Pentacam in random order. Intrasession repeatability and agreement were determined.Results: Employing WAM700, intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth was good (ICCs ≥ 0.992; CV 0.48-0.98%), yet worse than the values obtained for the Pentacam (ICCs ≥ 0.998; CV 0-0.33%). WAM700 showed excellent intrasession repeatability when used to measure the anterior chamber depth (Sw 0.03 mm). However, the repeatability of this device was inferior for central corneal thickness (Sw 4.24 μm) and keratometry measurements (Sw < 0.21 D) and was poor for corneal eccentricity (Sw 0.07; ICC 0.908; CV 14.58%). Agreement between WAM700 and Pentacam showed a high ICC for the keratometry measurements, central corneal thickness and anterior chamber depth (>0.972) but lower for corneal eccentricity (ICC 0.762).Conclusions: In healthy eyes, the WAM700 multidiagnostic device showed good intrasession repeatability for keratometry, central corneal thickness and anterior chamber depth measurements. Agreement between WAM700 and Pentacam was good for the anterior chamber depth measurement. However, these instruments cannot be considered interchangeable for keratometry, central corneal thickness and eccentricity readings. PB Taylor & Francis SN 0816-4622 YR 2024 FD 2024-03-17 LK https://hdl.handle.net/20.500.14352/104437 UL https://hdl.handle.net/20.500.14352/104437 LA eng NO Barrio AR, Antona B, Morago-Villanueva S, Martínez-Arribas V, González-Pérez M. Intrasession repeatability and agreement of the anterior corneal assessment provided by a multidiagnostic device. Clin Exp Optom. 2024 Mar 17:1-7. doi: 10.1080/08164622.2024.2324978. Epub ahead of print. PMID: 38494447. NO Received 07 Oct 2023, Accepted 23 Feb 2024, Published online: 17 Mar 2024 DS Docta Complutense RD 21 jul 2024