RT Journal Article T1 Comparison of cell mechanical measurements provided by Atomic ForceMicroscopy (AFM) and Micropipette Aspiration (MPA) A1 Daza, Rafael A1 González Bermúdez, Blanca A1 Cruces, Julia A1 Fuente del Rey, Mónica de la A1 Plaza, Gustavo R. A1 Arroyo Hernández, María A1 Elices Calafat, Manuel A1 Pérez Rigueiro, José A1 Guinea, Gustavo V. AB A comparative analysis of T-lymphocyte mechanical data obtained from Micropipette Aspiration (MPA) andAtomic Force Microscopy (AFM) is presented. Results obtained byfitting the experimental data to simple Hertzand Theret models led to non-Gaussian distributions and significantly different values of the elastic moduliobtained by both techniques. The use of more refined models, taking into account thefinite size of cells (sim-plified double contact and Zhou models) reduces the differences in the values calculated for the elastic moduli.Several possible sources for the discrepancy between the techniques are considered. The analysis suggests thatthe local nature of AFM measurements compared with the more general character of MPA measurementsprobably contributed to the differences observed. PB Elsevier SN 1878-0180 YR 2019 FD 2019-04-08 LK https://hdl.handle.net/20.500.14352/13698 UL https://hdl.handle.net/20.500.14352/13698 LA eng NO Ministerio de Economía y Competitividad (MINECO) NO Comunidad de Madrid DS Docta Complutense RD 5 abr 2025