Soria García, ÁngelaDel Hoyo Muñoz, JesúsSánchez Brea, Luis MiguelPastor Villarrubia, VerónicaGonzález Fernández, VerónicaElshorbagy, Mahmoud H.Alda Serrano, Javier2023-10-022023-10-022023-05-3110.1117/12.2664854https://hdl.handle.net/20.500.14352/88089We designed, fabricated and tested a Vector Diffractive Optical Element (VDOE) to simultaneously determine the Stokes vector of light. It comprises several sectors. Each one is a vector Fresnel zone plate which focuses the light on separate foci and has different polarization properties. The polarization state is calculated from their intensities. From simulations, we could identify the error sources that were analytically removed. The residual uncertainty after applying our corrections was as low as 6x10^(-5). The uncertainty obtained for our fabricated VDOE, 3.33 %, is competitive with the results from state-of-the-art techniques.engAttribution-NonCommercial-NoDerivatives 4.0 InternationalStokes polarimeter using vector diffractive optical elementsconference paperhttps://doi.org/10.1117/12.2664854open access535Vector Diffractive Optical Element, Polarimetry, Stokes parametersÓptica (Física)2209 Óptica