Azabal Arroyo, MagdalenaGarcia Otero, DMacorra García, José Carlos De La2023-06-202023-06-202004https://hdl.handle.net/20.500.14352/52242Aim: to study the effectiveness of an electronic apex locator (Justy II) in locating simulated horizontal and vertical fractures in single roots. Methods: an electronic apex locator (EAL) (Justy II, Yoshida Dentcraft, Tokyo, Japan) was used to measure the distance within the canal of horizontal (n=31) and vertical (n=31) fractures, created with a disk in single-rooted teeth. Accuracy of the EAL was evaluated by comparing the measurements with those made using a size 10 file. Data were analyzed with the non-parametric Passing and Bablok method. Results: for simulated horizontal fractures the EAL measured exactly the same length as a size 10 file, without constant or proportional errors. In vertical simulated fractures the EAL measured (on average) with a constant error of 7.5 mm shorter than the size 10 file; the difference had a wide confidence interval (–72.3 to 2.6 mm). Conclusion: in this laboratory study, the Justy II EAL was able to determine accurately the position of simulated horizontal fractures but was unreliable when measuring simulated vertical fractures.spaAccuracy of the Justy II apex locator in determining working length in simulated horizontal and vertical fracturesjournal articleopen accessLocalizador electrónico de ápicesFractura radicular horizontalFractura radicular verticalEndodonciaEndodoncia3213.13 Ortodoncia-Estomatología