Alfonso, José F.Fernández Vega Cueto, LuísLisa, CarlosAlfonso Bartolozzi, BelénPalacios, AnaMadrid Costa, David2024-01-112024-01-112023Alfonso JF, Fernández-Vega-Cueto L, Lisa C, Alfonso-Bartolozzi B, Palacios A, Madrid-Costa D. Clinical and Aberrometric Outcomes of a New Implantable Collamer Lens for Myopia and Presbyopia Correction in Phakic Patients. Journal of Refractive Surgery. 2023;39(9):589-596. doi:10.3928/1081597X-20230726-021081-597X10.3928/1081597X-20230726-02https://hdl.handle.net/20.500.14352/92448PURPOSE:To assess the clinical and aberrometric outcomes of a new Implantable Collamer Lens (EVO Viva ICL; STAAR Surgical) to correct moderate to high myopia and presbyopia. METHODS:The study included 80 eyes of 40 patients who had bilateral EVO Viva ICL implantation. Uncorrected (UDVA) and corrected (CDVA) distance visual acuities, refraction, presbyopic add power, binocular through-focus visual acuity, total ocular spherical aberration (SA), coma aberration, and root mean square of ocular higher order aberrations (RMS HOAs) (i-Trace Aberrometer; Tracey Technologies) for a 4.5-mm pupil size were evaluated. RESULTS:The mean binocular postoperative UDVA and CDVA were 0.09 ± 0.19 and 0.02 ± 0.03 logMAR, respectively. The postoperative spherical equivalent was −0.61 ± 0.54 diopters (D). The presbyopic add power reduced from +1.31 ± 0.74 D preoperatively to +0.44 ± 0.58 D after surgery (P < .0001). The mean visual acuity was 0.1 logMAR or better (20/25 or better) across the vergence range from +0.50 to −1.50 D, better than 0.2 logMAR (20/32 or better) up to the vergence of −2.00 D, and remained better than 0.3 logMAR (20/40 or better) up to the vergence of −2.50 D. The total ocular aberrations induced by EVO Viva ICL were −0.34 ± 0.09 µm of SA, 0.24 ± 0.18 µm of coma, and 0.26 ± 0.12 µm of RMS HOAs. CONCLUSIONS:The outcomes support that the new ICL might be a good alternative for myopia and presbyopia correction in patients aged between 45 and 55 years. Further studies are needed to evaluate the threshold lens misalignment from which the patient´s visual quality would be affected.engClinical and Aberrometric Outcomes of a New Implantable Collamer Lens for Myopia and Presbyopia Correction in Phakic Patientsjournal article1938-2391https://doi.org/10.3928/1081597X-20230726-02https://journals.healio.com/doi/full/10.3928/1081597X-20230726-02restricted access617.7Óptica y optometría2209.15 Optometría