Gómez Pedrero, José AntonioAlbarrán Diego, CésarGarcía Montero, MaríaGarzón Jiménez, NuriaGonzález Fernández, Verónica2023-11-152023-11-152023-08-31Gómez Pedrero, J. A., Albarrán Diego, C., García Montero, M. et al. «Influence of Instrumental Factors in the Measurement of Power Profiles of Intraocular Lenses with a Commercial Deflectometer». Applied Sciences, vol. 13, n.o 17, agosto de 2023, p. 9882. DOI.org (Crossref), https://doi.org/10.3390/app13179882.10.3390/app13179882https://hdl.handle.net/20.500.14352/88732Deflectometry is an optical technique for determining properties such as power distribution, wavefront, etc., and measurement of the optical properties of an intraocular lens can provide relevant information for clinicians. The aim of the current study was to establish a protocol for measuring lens power maps and profiles of various optical designs of intraocular lenses with a deflectometer based on the phase-shifting Schlieren principle (NIMO TR1504, Lambda-X, Nivelles, Belgium). The results are discussed with respect to accuracy and repeatability, the influence of the use of filters, and whether to consider the intraocular lens as a thin or thick lens.engAttribution 4.0 InternationalInfluence of Instrumental Factors in the Measurement of Power Profiles of Intraocular Lenses with a Commercial Deflectometerjournal article2076-3417https://doi.org/10.3390/app13179882open access617.7-089.243681.73617.75Intraocular lensDeflectometryPower profileÓptica y optometríaÓptica geométrica e instrumental2209 Óptica3311.09 Lentes2209.15 Optometría