Clinical evaluation of the ADVIA Centaur XPT chemiluminescent immunoassay for equine insulin measurement

dc.contributor.authorRey Conejo, R.
dc.contributor.authorToribio, R.E.
dc.contributor.authorMoller, S.
dc.contributor.authorMüller, E.
dc.contributor.authorFores Jackson, Paloma
dc.date.accessioned2025-08-18T07:34:33Z
dc.date.available2025-08-18T07:34:33Z
dc.date.issued2025
dc.descriptionCredit authorship contribution statement: R. Rey-Conejo: Writing – original draft, Methodology, Investigation, Formal analysis, Conceptualization. R.E. Toribio: Writing – review & editing, Supervision, Conceptualization. S. Moller: ¨ Writing – review & editing, Resources. E. Müller: Writing – review & editing, Resources. P. Fores-Jackson: Writing – review & editing, Supervision, Conceptualization
dc.description.abstractBackground: Insulin dysregulation is a key component of equine metabolic syndrome and is commonly assessed using basal insulin concentrations. Different analyzers may yield variable insulin results, limiting comparability. Aims/Objectives: To assess the analytical performance of the ADVIA Centaur XPT chemiluminescent immunoassay (CLIA) for equine insulin measurement and to establish adapted decision thresholds based on comparison with a previously used CLIA method. Methods: Precision, linearity, and dilution recovery were assessed for the ADVIA Centaur XPT. A total of 89 equine serum samples submitted for basal insulin testing were measured using the ADVIA Centaur XPT and Immulite 2000 XPi analyzers. Results were compared using Deming regression, Bland-Altman analysis, and weighted kappa statistics to evaluate agreement in classification. Results: Intra-assay coefficients of variation were 3.56 %, 2.01 %, and 1.92 % for low, medium, and high insulin concentrations, respectively; inter-assay variation was 5.19 %, 5.78 %, and 5.68 %. Deming regression showed a proportional bias, with the ADVIA Centaur XPT consistently measuring lower insulin concentrations compared to the Immulite 2000 XPi. Bland-Altman analysis confirmed this bias across the measurement range. Spearman correlation between the two methods was 0.92 (95 % CI: 0.87–0.94; P < 0.0001), indicating a strong association in rank order. Classification agreement using adapted decision limits yielded a weighted kappa of 0.82, indicating strong agreement. Conclusion: The ADVIA Centaur XPT demonstrated acceptable precision. When insulin concentrations were classified using adapted thresholds, the agreement between the two CLIA analyzers was strong, suggesting that the ADVIA Centaur XPT may be suitable for measurement of equine insulin concentration.
dc.description.departmentDepto. de Medicina y Cirugía Animal
dc.description.facultyFac. de Veterinaria
dc.description.refereedTRUE
dc.description.statuspub
dc.identifier.citationRey-Conejo, R., Toribio, R. E., Möller, S., Müller, E., & Fores-Jackson, P. (2025). Clinical evaluation of the ADVIA Centaur XPT chemiluminescent immunoassay for equine insulin measurement. Journal of equine veterinary science, 152, 105661. https://doi.org/10.1016/j.jevs.2025.105661
dc.identifier.doi10.1016/j.jevs.2025.105661
dc.identifier.essn1542-7412
dc.identifier.issn0737-0806
dc.identifier.officialurlhttps://doi.org/10.1016/j.jevs.2025.105661
dc.identifier.pmid40752549
dc.identifier.relatedurlhttps://www.sciencedirect.com/science/article/pii/S0737080625003193
dc.identifier.urihttps://hdl.handle.net/20.500.14352/123166
dc.issue.number105661
dc.journal.titleJournal of Equine Veterinary Science
dc.language.isoeng
dc.page.final6
dc.page.initial1
dc.publisherElsevier
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internationalen
dc.rights.accessRightsopen access
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subject.cdu619: 616. 4
dc.subject.keywordADVIA centaur XPT
dc.subject.keywordChemiluminescent immunosassay
dc.subject.keywordEquine insulin
dc.subject.keywordImmulite 2000 XPi
dc.subject.keywordInsulin dysregulation
dc.subject.ucmGanado equino
dc.subject.ucmEndocrinología
dc.subject.unesco3205.02 Endocrinología
dc.titleClinical evaluation of the ADVIA Centaur XPT chemiluminescent immunoassay for equine insulin measurement
dc.typejournal article
dc.type.hasVersionVoR
dc.volume.number152
dspace.entity.typePublication
relation.isAuthorOfPublication96867e97-ddb4-4f8c-9d6d-ad9fee630eba
relation.isAuthorOfPublication.latestForDiscovery96867e97-ddb4-4f8c-9d6d-ad9fee630eba

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
1-s2.0-S0737080625003193-main.pdf
Size:
864.14 KB
Format:
Adobe Portable Document Format

Collections