Aviso: para depositar documentos, por favor, inicia sesión e identifícate con tu cuenta de correo institucional de la UCM con el botón MI CUENTA UCM. No emplees la opción AUTENTICACIÓN CON CONTRASEÑA
 

Differences in Cyclic Fatigue Resistance at Apical and Coronal Levels of Reciproc and WaveOne New Files

dc.contributor.authorArias Paniagua, Ana María
dc.contributor.authorPérez-Higueras Sánchez-Escalonilla, Juan
dc.contributor.authorMacorra García, José Carlos De La
dc.date.accessioned2023-06-20T03:37:09Z
dc.date.available2023-06-20T03:37:09Z
dc.date.issued2012
dc.description.abstractIntroduction: Recent improvements in alloys, kinematics, and concepts have been combined to increase the cyclic fatigue resistance of nickel-titanium (NiTi) instruments. The aim of this study was to compare the cyclic fatigue resistance of new M-Wire reciprocating WaveOne (Dentsply Maillefer, Ballaigues, Switzerland) and Reciproc (VDW GmbH, Munich, Germany) files at 2 levels. Methods: Sixty Reciproc and 60 WaveOne new files were fixed to a specifically designed device and tested in tempered steel canals with a 3-mm radius and a 60° angle of curvature. The motor used was programmed as defined by each manufacturer, and the specific reciprocating motion was followed. Thirty files of each brand were tested at 5 mm, and 30 were tested at 13 mm from their tips. The time to failure was registered. Weibull analysis was used to calculate the mean life, beta, and eta parameters. Results: The probability of the mean life was higher for Reciproc than WaveOne files at both levels, with the probability of the Reciproc mean life being 62% higher than that of WaveOne at 5 mm from the tip and 100% higher at 13 mm (all statistically significant). The probability of the mean life was higher at 5 mm than at 13 mm in both systems. The results may have been partially conditioned by the different motions that manufacturers propose for each system. Conclusions: Reciproc files were more resistant to cyclic fatigue than WaveOne files at both distances from the tip. Both systems had greater cyclic fatigue resistance at 5 mm than at 13 mm from the tip.en
dc.description.departmentDepto. de Especialidades Clínicas Odontológicas
dc.description.facultyFac. de Odontología
dc.description.refereedTRUE
dc.description.statuspub
dc.eprint.idhttps://eprints.ucm.es/id/eprint/24088
dc.identifier.citationArias, A.; Perez-Higueras, J. J.; De La Macorra, J. C. Differences in Cyclic Fatigue Resistance at Apical and Coronal Levels of Reciproc and WaveOne New Files. Journal of Endodontics 2012, 38(9), 1244–1248. doi:10.1016/j.joen.2012.05.022.
dc.identifier.doi10.1016/j.joen.2012.05.022
dc.identifier.issn0099-2399
dc.identifier.officialurlhttps//doi.org/10.1016/j.joen.2012.05.022
dc.identifier.urihttps://hdl.handle.net/20.500.14352/44059
dc.issue.number9
dc.journal.titleJournal of Endodontics
dc.language.isospa
dc.page.final1248
dc.page.initial1244
dc.publisherElsevier
dc.rights.accessRightsopen access
dc.subject.ucmAparatos y material odontológicos
dc.subject.ucmMateriales dentales
dc.subject.ucmEndodoncia
dc.subject.unesco3311.03 Instrumentos Para Odontología
dc.subject.unesco3213.13 Ortodoncia-Estomatología
dc.subject.unesco3213.13 Ortodoncia-Estomatología
dc.titleDifferences in Cyclic Fatigue Resistance at Apical and Coronal Levels of Reciproc and WaveOne New Filesen
dc.typejournal article
dc.volume.number38
dspace.entity.typePublication
relation.isAuthorOfPublicationf4270363-d6c5-4ffc-9ebe-c90482776ec1
relation.isAuthorOfPublication2dbd67ad-7fd0-4ba6-b60f-a1f7923a9e01
relation.isAuthorOfPublication0b38ba8d-c9aa-404d-8ec6-b8c2d6a78884
relation.isAuthorOfPublication.latestForDiscoveryf4270363-d6c5-4ffc-9ebe-c90482776ec1

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Differences_in_Cyclic_Fatigue_Resistance_at_Apical_and_Coronal_Levels_of_Reciproc_and_WaveOne_New_Files.pdf
Size:
452.39 KB
Format:
Adobe Portable Document Format

Collections