Validation of a rapid and easy-to-apply method to simultaneously quantify co-loaded dexamethasone and melatonin plgamicrospheres by hplc-uv: encapsulation efficiency and in vitro release

dc.contributor.authorVicario De La Torre, Marta
dc.contributor.authorHerrero Vanrell, María Del Rocío
dc.contributor.authorAndrés Guerrero, Vanesa
dc.contributor.authorBravo Osuna, Irene
dc.contributor.authorMolina Martínez, Irene Teresa
dc.contributor.authorBrugnera, Marco
dc.date.accessioned2025-01-22T11:29:22Z
dc.date.available2025-01-22T11:29:22Z
dc.date.issued2022-01-26
dc.description.abstractThis paper discusses the development and validation of a rapid method for the reversed phase HPLC-UV quantification of biodegradable poly(D,L-lactic-co-glycolic) acid (PLGA) micro-spheres co-loaded with two neuroprotective agents (dexamethasone and melatonin) (DX-MEL-MSs) to be intravitreally administered as a promising glaucoma treatment. The study was performed to validate two procedures that quantify the content of the two active substances entrapped into the polymer matrix during an encapsulation efficiency assay and the amount of drugs liberated over time during the in vitro release assay. The reversed-phase method allowed for the simultaneous determination of dexamethasone and melatonin, which were respectively detected at 240.5 and 222.7 nm. Chromatographic separation was performed using an Ascentis® C18 HPLC Column (25 cm × 4.6 mm, 5 μm) with an isocratic mobile phase composed of methanol-water (70:30, v/v) with 1.0 mL min−1 flow rate. The two procedures were validated analytically in terms of system suitability testing, specificity, linearity, precision, accuracy, sensitivity, and robustness. Both the validated procedures were applied to characterize DX-MEL-MSs and were found appropriate to quantify the drug quantities encapsulated and estimate their release profile over 10 days. The validation study designed in this work can be helpful for planning any other protocols that refer to the quantification of PLGA based drug delivery systems.
dc.description.departmentDepto. de Farmacia Galénica y Tecnología Alimentaria
dc.description.facultyFac. de Farmacia
dc.description.refereedTRUE
dc.description.sponsorshipUniversidad Complutense de Madrid. Research Group UCM 920415.
dc.description.sponsorshipHorizon 2020 research and innovation programme (European Commission)
dc.description.statuspub
dc.identifier.citationBrugnera, M.; Vicario-de-la- Torre, M.; Andrés-Guerrero, V.; Bravo-Osuna, I.; Molina-Martínez, I.T.; Herrero-Vanrell, R. Validation of a Rapid and Easy-to-Apply Method to Simultaneously Quantify Co-Loaded Dexamethasone and Melatonin PLGA Microspheres by HPLC-UV: Encapsulation Efficiency and In Vitro Release. Pharmaceutics 2022, 14, 288. https://doi.org/ 10.3390/pharmaceutics14020288
dc.identifier.doi10.3390/pharmaceutics14020288
dc.identifier.officialurlhttps://doi.org/10.3390/pharmaceutics14020288
dc.identifier.urihttps://hdl.handle.net/20.500.14352/115544
dc.issue.number2
dc.journal.titlePharmaceutics
dc.language.isoeng
dc.publisherMDPI
dc.relation.projectID81344
dc.rightsAttribution 4.0 Internationalen
dc.rights.accessRightsopen access
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/
dc.subject.cdu615.4
dc.subject.cdu617.7
dc.subject.keywordGlaucoma
dc.subject.keywordMelatonin
dc.subject.keywordDexamethasone
dc.subject.keywordValidation
dc.subject.keywordHPLC-UV
dc.subject.keywordPLGA
dc.subject.keywordMicrospheres
dc.subject.keywordCo-delivery
dc.subject.keywordEncapsulation efficiency
dc.subject.keywordIn vitro release
dc.subject.ucmTecnología farmaceútica
dc.subject.ucmOftalmología
dc.subject.unesco3201.09 Oftalmología
dc.subject.unesco2403 Bioquímica
dc.titleValidation of a rapid and easy-to-apply method to simultaneously quantify co-loaded dexamethasone and melatonin plgamicrospheres by hplc-uv: encapsulation efficiency and in vitro release
dc.typejournal article
dc.type.hasVersionVoR
dc.volume.number14
dspace.entity.typePublication
relation.isAuthorOfPublication765de4d4-2555-427f-ae06-236d9ecbc05b
relation.isAuthorOfPublicatione19672b5-d6f7-400a-b591-b903bc396955
relation.isAuthorOfPublication14096cb3-8372-40d2-8fe1-79d0c4c1ddd7
relation.isAuthorOfPublication491381a5-fc5a-4cd5-a0b0-9b6cfdae6706
relation.isAuthorOfPublication43cca2c8-2da7-45dd-bce6-4b70a86a3877
relation.isAuthorOfPublication.latestForDiscovery765de4d4-2555-427f-ae06-236d9ecbc05b

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Validation of a Rapid and Easy-to-Apply Method to Simultaneously Quantify Co-Loaded Dexamethasone_final.pdf
Size:
4.69 MB
Format:
Adobe Portable Document Format

Collections