Electrochemical immunosensor for rapid and sensitive determination of estradiol

dc.contributor.authorOjeda Fernández, Irene
dc.contributor.authorLopez-Montero, Judith
dc.contributor.authorMoreno Guzmán, María
dc.contributor.authorJanegitz, Bruno
dc.contributor.authorGonzález Cortés, Araceli
dc.contributor.authorYáñez-Sedeño Orive, Paloma
dc.contributor.authorPingarrón Carrazón, José Manuel
dc.date.accessioned2024-01-09T08:39:39Z
dc.date.available2024-01-09T08:39:39Z
dc.date.issued2012
dc.description.abstractThis work describes the preparation of an electrochemical immunosensor for estradiol based on the surface modification of a screen printed carbon electrode with grafted p-aminobenzoic acid followed by covalent binding of streptavidin (Strept) and immobilization of biotinylated anti-estradiol (antiestradiol-Biotin). The hormone determination was performed by applying a competitive immunoassay with peroxidase-labelled estradiol (HRP–estradiol) and measurement of the amperometric response at −200 mV using hydroquinone (HQ) as redox mediator. The calibration curve for estradiol exhibited a linear range between 1 and 250 pg mL−1 (r = 0.990) and a detection limit of 0.77 pg mL−1 was achieved. Cross-reactivity studies with other hormones related with estradiol at physiological concentration levels revealed the practical specificity of the developed method for estradiol. A good reproducibility, with RSD = 5.9% (n = 8) was also observed. The operating stability of a single bioelectrode modified with anti-estradiol-Biotin-Strept was nine days when it was stored at 8 ◦C under humid conditions between measurements. The developed immunosensor was applied to the analysis of certified serum and spiked urine samples with good results
dc.description.departmentDepto. de Química Analítica
dc.description.facultyFac. de Ciencias Químicas
dc.description.refereedTRUE
dc.description.sponsorshipMinisterio de Ciencia e Innovación (España)
dc.description.sponsorshipComunidad de Madrid
dc.description.statuspub
dc.identifier.citationOjeda, I., et al. «Electrochemical Immunosensor for Rapid and Sensitive Determination of Estradiol». Analytica Chimica Acta, vol. 743, septiembre de 2012, pp. 117-24. https://doi.org/10.1016/j.aca.2012.07.002.
dc.identifier.doi10.1016/j.aca.2012.07.002
dc.identifier.issn0003-2670
dc.identifier.officialurlhttps://doi.org/10.1016/j.aca.2012.07.002
dc.identifier.urihttps://hdl.handle.net/20.500.14352/91916
dc.journal.titleAnalytica Chimica Acta
dc.language.isoeng
dc.page.final124
dc.page.initial117
dc.publisherElsevier
dc.relation.projectID(Projects CTQ2009-12650)
dc.relation.projectIDS2009/PPQ1642
dc.rights.accessRightsrestricted access
dc.subject.cdu543
dc.subject.keywordEstradiol
dc.subject.keywordElectrochemical immunosensor
dc.subject.keywordScreen-printed electrodes
dc.subject.keywordSerum
dc.subject.keywordUrine
dc.subject.ucmQuímica analítica (Química)
dc.subject.unesco2301 Química Analítica
dc.titleElectrochemical immunosensor for rapid and sensitive determination of estradiol
dc.typejournal article
dc.type.hasVersionVoR
dc.volume.number743
dspace.entity.typePublication
relation.isAuthorOfPublicationb4631aae-bbc4-470a-922b-73f321e2515a
relation.isAuthorOfPublication96968530-a80d-4987-a341-f4a96697f66c
relation.isAuthorOfPublication8a730734-cdf5-4b59-b017-12e2320c872e
relation.isAuthorOfPublication06ab2fa0-2b2d-4b58-80ed-82f9cc767175
relation.isAuthorOfPublication8808f99c-5f56-4562-839a-517626c76dad
relation.isAuthorOfPublication.latestForDiscoveryb4631aae-bbc4-470a-922b-73f321e2515a

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Determination_of_estradiol.pdf
Size:
805.06 KB
Format:
Adobe Portable Document Format

Collections