Holographic Fabry-Perot spectrometer
dc.contributor.author | Martínez Matos, Óscar | |
dc.contributor.author | Rodrigo Martín-Romo, José Augusto | |
dc.contributor.author | Vaveliuk, Pablo | |
dc.contributor.author | Calvo Padilla, María Luisa | |
dc.date.accessioned | 2023-06-20T03:38:58Z | |
dc.date.available | 2023-06-20T03:38:58Z | |
dc.date.issued | 2011-02-15 | |
dc.description | © 2011 Optical Society of America. We thank A. Cámara Iglesias for valuable discussions and advice. Financial support from the Spanish Ministry of Science and Innovation under project TEC 2008-04105 is acknowledged. P.Vaveliuk acknowledges Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq), Brazil, for financial support. | |
dc.description.abstract | We propose a spectrum analyzer based on the properties of a hologram recorded with the field transmitted by a Fabry-Perot etalon. The spectral response of this holographic Fabry-Perot spectrometer (HFPS) is analytically investigated in the paraxial approximation and compared with a conventional Fabry-Perot etalon of similar characteristics. We demonstrate that the resolving power is twice increased and the free spectral range (FSR) is reduced to one-half. The proposed spectrometer could improve the operational performance of the etalon because it can exhibit high efficiency and it would be insensible to environmental conditions such as temperature and vibrations. Our analysis also extends to another variant of the HFPS based on holographic multiplexing of the transmitted field of a Fabry-Perot etalon. This device increases the FSR, keeping the same HFPS performance. | |
dc.description.department | Depto. de Óptica | |
dc.description.faculty | Fac. de Ciencias Físicas | |
dc.description.refereed | TRUE | |
dc.description.sponsorship | Ministerio de Ciencia e Innovación, España | |
dc.description.sponsorship | Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq), Brazil | |
dc.description.status | pub | |
dc.eprint.id | https://eprints.ucm.es/id/eprint/25137 | |
dc.identifier.doi | 10.1364/OL.36.000564 | |
dc.identifier.issn | 0146-9592 | |
dc.identifier.officialurl | http://dx.doi.org/10.1364/OL.36.000564 | |
dc.identifier.relatedurl | http://www.opticsinfobase.org | |
dc.identifier.uri | https://hdl.handle.net/20.500.14352/44154 | |
dc.issue.number | 4 | |
dc.journal.title | Optics Letters | |
dc.language.iso | eng | |
dc.page.final | 566 | |
dc.page.initial | 564 | |
dc.publisher | Optical Society of America | |
dc.relation.projectID | TEC 2008-04105 | |
dc.rights.accessRights | open access | |
dc.subject.cdu | 535 | |
dc.subject.keyword | Optics | |
dc.subject.ucm | Óptica (Física) | |
dc.subject.unesco | 2209.19 Óptica Física | |
dc.title | Holographic Fabry-Perot spectrometer | |
dc.type | journal article | |
dc.volume.number | 36 | |
dcterms.references | 1. M. Born and E. Wolf, Principles of Optics (Cambridge University, 1999). 2. J. M. Vaughan, The Fabry–Perot Interferometer: History, Theory, Practice and Applications (Taylor & Francis, 1989). 3. Ch. Fabry and A. Perot, Ann. Chim. Phys. 7, 459 (1897). 4. L. Solymar and D. J. Cooke, Volume Holography and Volume Gratings (Academic, London, 1981). 5. O. Martinez-Matos, J. A. Rodrigo, M. P. Hernández-Garay, J. G. Izquierdo, R. Weigand, M. L. Calvo, P. Cheben, P. Vaveliuk, and L. Bañares, Opt. Lett. 35, 652 (2010). 6. F. del Monte, O. Martínez-Matos, J. A. Rodrigo, M. L. Calvo, and P. Cheben, Adv. Mater. 18, 2014 (2006). | |
dspace.entity.type | Publication | |
relation.isAuthorOfPublication | b6643c3d-f635-48d3-a642-922a4b2e595c | |
relation.isAuthorOfPublication | e2846481-608d-43dd-a835-d70f73a4dd48 | |
relation.isAuthorOfPublication.latestForDiscovery | b6643c3d-f635-48d3-a642-922a4b2e595c |