High-frequency ultrasound biomicroscopy of silicone posterior chamber phakic intraocular lens for hyperopia

dc.contributor.authorGarcía Feijoo, Julián
dc.contributor.authorHernández Matamoros, J.L
dc.contributor.authorCastillo Gómez, Alfredo
dc.contributor.authorLázaro, C
dc.contributor.authorMéndez Hernández, Carmen Dora
dc.contributor.authorMartín, T
dc.contributor.authorMartínez De La Casa Fernández-Borrella, José María
dc.contributor.authorGarcía Sánchez, Julián
dc.date.accessioned2024-08-02T08:57:56Z
dc.date.available2024-08-02T08:57:56Z
dc.date.issued2003-10-01
dc.description.abstractPurpose: To study the intraocular position and anatomic relationships of the PRL-III phakic refractive lens (PRL), a posterior chamber phakic intraocular lens (PCP IOL), in cases of hyperopia using ultrasound biomicroscopy (UBM). Setting: Centro Oftalmológico Real Vision, Madrid, Spain, and Instituto de Investigaciones Oftalmológicas Ramón Castroviejo, Universidad Complutense, Madrid, Spain. Methods: Eleven phakic hyperopic eyes of 6 patients who had PRL implantation were examined by UBM 1 month after surgery. The PRL position, PRL-crystalline lens peripheral distance, and central distance between the corneal endothelium and the PRL were measured. Results: Eight eyes had both haptics on the zonule, 2 had 1 haptic in the sulcus and 1 on the zonule, and 1 had 1 haptic in the sulcus and the other in the ciliary body. The mean PCP IOL-crystalline lens peripheral distance in the minor axis was 239.7 μm ± 179.4 (SD) and the mean PCP IOL-endothelium central distance, 2146.98 ± 219.6 μm. Contact between the PCP IOL and crystalline lens was observed in 1 eye. Conclusions: In this study of hyperopic eyes, the PRL was located on the zonule in most cases. However, the location of the haptics in the sulcus and contact between the PCP IOL and the crystalline lens that occurred in some cases suggest further study of possible long-term complications is needed. © 2003 ASCRS and ESCRS.
dc.description.departmentDepto. de Inmunología, Oftalmología y ORL
dc.description.facultyFac. de Medicina
dc.description.refereedTRUE
dc.description.statuspub
dc.identifier.citationGarcía-Feijoó J, Hernández-Matamoros JL, Castillo-Gómez A, Lázaro C, Méndez-Hernández C, Martín T, Martínez de la Casa JM, García-Sánchez J. High-frequency ultrasound biomicroscopy of silicone posterior chamber phakic intraocular lens for hyperopia. J Cataract Refract Surg. 2003 Oct;29(10):1940-6. doi: 10.1016/s0886-3350(03)00240-2. PMID: 14604714.
dc.identifier.doi10.1016/S0886-3350(03)00240-2
dc.identifier.officialurlhttps://doi.org/10.1016/S0886-3350(03)00240-2
dc.identifier.relatedurlhttps://journals.lww.com/jcrs/abstract/2003/10000/high_frequency_ultrasound_biomicroscopy_of.30.aspx
dc.identifier.urihttps://hdl.handle.net/20.500.14352/107355
dc.issue.number10
dc.journal.titleJournal of Cataract and Refractive Surgery
dc.language.isoeng
dc.page.final1946
dc.page.initial1940
dc.publisherLippincott, Williams & Wilkins
dc.rights.accessRightsrestricted access
dc.subject.cdu616.75-089.243
dc.subject.ucmOftalmología
dc.subject.unesco3201.09 Oftalmología
dc.titleHigh-frequency ultrasound biomicroscopy of silicone posterior chamber phakic intraocular lens for hyperopia
dc.typejournal article
dc.type.hasVersionVoR
dc.volume.number29
dspace.entity.typePublication
relation.isAuthorOfPublication558b8023-6d72-4dff-9f99-2e60f6f31843
relation.isAuthorOfPublicatione025c156-d03e-48a0-a6a0-c16196262baf
relation.isAuthorOfPublication533dd90f-2b34-4a5f-9cf8-ab8d19454edd
relation.isAuthorOfPublication273a99c3-2c9f-4dd0-8939-b7ff3593124c
relation.isAuthorOfPublicationeff24faf-e089-4604-906d-bbd7f42be7f1
relation.isAuthorOfPublication.latestForDiscovery558b8023-6d72-4dff-9f99-2e60f6f31843

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Ultrasound Biomicroscopy of Phakic IOL for Hyperopia - García-Feijoó et al. 2003.pdf
Size:
199.93 KB
Format:
Adobe Portable Document Format

Collections