Non-invasive blood pressure measurement in conscious rabbits: A comparison of Doppler ultrasonic and oscillometric devices

dc.contributor.authorGonzález González, Enrique
dc.contributor.authorGonzález Alonso-Alegre, Elisa-Margarita
dc.contributor.authorMontesinos Barceló, Andrés
dc.contributor.authorCaro Vadillo, Alicia
dc.date.accessioned2024-09-30T12:48:52Z
dc.date.available2024-09-30T12:48:52Z
dc.date.issued2024
dc.description.abstractBackground This study aimed to compare Doppler (the most commonly used technique for non-invasive indirect blood pressure [NIBP] measurement in rabbits) and oscillometric devices for NIBP measurements in conscious rabbits. Methods Blood pressure (BP) measurements were obtained for 49 conscious rabbits using the Doppler and oscillometric devices. Each measurement was repeated in triplicate. Three rabbits were excluded from the study because it was not possible to obtain BP measurements using the oscillometric device. The American College of Veterinary Internal Medicine (ACVIM) guidelines were followed to compare the results obtained with the two devices. Results A total of 164 systolic BP measurements were obtained with the Doppler device and 182 were obtained with the oscillometric technique. The mean of the three values obtained for each animal with each device was used for statistical analysis. The analysis demonstrated a poor level of agreement between the measurements obtained with the two devices. Limitations Oscillometric devices rely on built-in software algorithms to perform calculations and produce their measurements. However, the oscillometric devices currently available only have the software for dogs and cats installed. Therefore, these conclusions should be revised when a device with software specifically for rabbits is commercialised. Conclusions This study indicates that oscillometric devices should not be used as a substitute for the Doppler technique when measuring BP in rabbits.
dc.description.departmentDepto. de Medicina y Cirugía Animal
dc.description.facultyFac. de Veterinaria
dc.description.refereedTRUE
dc.description.statuspub
dc.identifier.citationGonzález-González E, González-Alonso-Alegre E, Montesinos-Barceló A, Caro-Vadillo A. Non-invasive blood pressure measurement in conscious rabbits: A comparison of Doppler ultrasonic and oscillometric devices. Vet Rec. 2024;e4399. https://doi.org/10.1002/vetr.4399
dc.identifier.doi10.1002/vetr.4399
dc.identifier.essn2052-6113
dc.identifier.officialurlhttps://doi.org/10.1002/vetr.4399
dc.identifier.pmid39267367
dc.identifier.urihttps://hdl.handle.net/20.500.14352/108469
dc.issue.numbere4399
dc.journal.titleVetRecord
dc.language.isoeng
dc.page.final9
dc.page.initial1
dc.publisherWiley
dc.rightsAttribution-NonCommercial 4.0 Internationalen
dc.rights.accessRightsopen access
dc.rights.urihttp://creativecommons.org/licenses/by-nc/4.0/
dc.subject.cdu636.09
dc.subject.keywordDoppler
dc.subject.keywordNon-invasive blood pressure measurement
dc.subject.keywordOscillometric technique
dc.subject.keywordRabbit medicine
dc.subject.ucmVeterinaria
dc.subject.unesco3109 Ciencias Veterinarias
dc.titleNon-invasive blood pressure measurement in conscious rabbits: A comparison of Doppler ultrasonic and oscillometric devices
dc.typejournal article
dc.type.hasVersionVoR
dc.volume.number2024
dspace.entity.typePublication
relation.isAuthorOfPublicationc0108e3e-acb9-4474-8888-469ac8b7c155
relation.isAuthorOfPublication4027b097-f516-461e-86a1-c5b3cbfdaa9b
relation.isAuthorOfPublication734157f1-d6f0-44a6-92d0-88fd129ca4fa
relation.isAuthorOfPublication8e6c38ec-45b7-40a0-bd0f-d3341387671c
relation.isAuthorOfPublication.latestForDiscoveryc0108e3e-acb9-4474-8888-469ac8b7c155

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Non‐invasive blood.pdf
Size:
518.95 KB
Format:
Adobe Portable Document Format

Collections