Squareness error calibration of a CMM for quality control of ophthalmic lenses

dc.contributor.authorQuiroga Mellado, Juan Antonio
dc.contributor.authorRodríguez Ibáñez, Diego
dc.contributor.authorAlonso Fernández, José
dc.date.accessioned2023-06-19T13:22:13Z
dc.date.available2023-06-19T13:22:13Z
dc.date.issued2013-09
dc.description© The Author(s) 2013. This article is published with open access at Springerlink.com. This work was supported by the Spanish Ministry of Science and Technology under grant DPI2009-09023. This article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
dc.description.abstractThis paper proposes a calibration procedure for the squareness error of small-sized coordinate measuring machines like those used for ophthalmic lenses quality control. The proposed procedure only requires a spherical lens as a reference surface and it only needs to be measured once making this calibration method faster and cheaper than the existing ones.
dc.description.departmentDepto. de Óptica
dc.description.facultyFac. de Ciencias Físicas
dc.description.refereedTRUE
dc.description.sponsorshipSpanish Ministry of Science and Technology
dc.description.statuspub
dc.eprint.idhttps://eprints.ucm.es/id/eprint/23293
dc.identifier.doi10.1007/s00170-013-4746-y
dc.identifier.issn0268-3768
dc.identifier.officialurlhttp://dx.doi.org/10.1007/s00170-013-4746-y
dc.identifier.relatedurlhttp://link.springer.com
dc.identifier.urihttps://hdl.handle.net/20.500.14352/33371
dc.issue.number1-4
dc.journal.titleInternational Journal of Advanced Manufacturing Technology
dc.language.isoeng
dc.page.final493
dc.page.initial487
dc.publisherSpringer London Ltd.
dc.relation.projectIDDPI2009-09023
dc.rightsAtribución 3.0 España
dc.rights.accessRightsopen access
dc.rights.urihttps://creativecommons.org/licenses/by/3.0/es/
dc.subject.cdu535
dc.subject.keywordCMM
dc.subject.keywordCalibration
dc.subject.keywordSquareness Error
dc.subject.keywordVolumetric Error
dc.subject.keywordOphthalmic Lenses Quality Control
dc.subject.ucmÓptica (Física)
dc.subject.unesco2209.19 Óptica Física
dc.titleSquareness error calibration of a CMM for quality control of ophthalmic lenses
dc.typejournal article
dc.volume.number68
dcterms.references1. Barakat NA, ElbestawiMA, Spence AD (2000) Kinematic and geometric error compensation of a coordinate measuring machine. Int J Mach Tools Manuf 40(6):833–850. 2. Okafor AC, Ertekin YM (2000) Derivation of machine tool error models and error compensation procedure for three axes vertical machining center using rigid body kinematics. Int J Mach Tools Manuf 40 (8): 1199–1213. 3. Ramesh R, Mannan MA, Poo AN (2000) Error compensation in machine tools—a review: part I: geometric, cutting-force induced and fixture-dependent errors. Int J Mach Tools Manuf 40(9): 1235–1256. 4. Wand BT, Bressler VE, Ruland R Squareness error mapping of a coordinate measuring machine using a two-dimensional reference object, SLAC Publication, SLAC-PUB-5836. 5. Lee ES, Burdekin M (2001) A hole-plate artifact design for the volumetric error calibration of CMM. Int J Adv Manuf Technol 17:508–515. doi:10.1007/s001700170151. 6. Liebrich T, Bringmann B, KnappW(2009) Calibration of a 3D-ball plate. Precision Engineering 33(1):1–6. 7. Kruth JP, Zhou L, Van den Bergh C, Vanherck P (2003) A method for squareness error verification on a coordinate measuring machine. Int J Adv Manuf Technol 21:874–878. doi:10.1007/s00170-002-1408-x. 8. Lagarias JC, Reeds JA, Wright MH,Wright PE (1998) Convergence properties of the nelder mead simplex method in low dimensions. SIAM J Optim 9(1):112–147.
dspace.entity.typePublication
relation.isAuthorOfPublication1c171089-8e25-448f-bcce-28d030f8f43a
relation.isAuthorOfPublicationf7b5b178-742c-418d-80d3-769a169f6dd9
relation.isAuthorOfPublication.latestForDiscovery1c171089-8e25-448f-bcce-28d030f8f43a

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
QuirogaJA102.pdf
Size:
728.86 KB
Format:
Adobe Portable Document Format

Collections