Evaluation of Rotation and Visual Outcomes After Implantation of Monofocal and Multifocal Toric Intraocular Lenses
dc.contributor.author | Garzón Jiménez, Nuria | |
dc.contributor.author | Poyales, Francisco | |
dc.contributor.author | Ortíz de Zárate, Begoña | |
dc.contributor.author | Ruiz-García, José Luis | |
dc.contributor.author | Quiroga Mellado, Juan Antonio | |
dc.date.accessioned | 2023-06-18T06:45:46Z | |
dc.date.available | 2023-06-18T06:45:46Z | |
dc.date.issued | 2015-03-12 | |
dc.description | Este documento es un postprint. | |
dc.description.abstract | PURPOSE: To evaluate rotational stability and its influ-ence on postoperative visual acuity of different monofocal and multifocal toric intraocular lenses (IOLs). METHODS: A prospective interventional study was designed. Ninety-one patients with a mean age of 71.65 ± 11.82 years were implanted with toric IOLs after phacoemulsification. Three monofocal toric IOLs (the Lentis LT [Oculentis, Berlin, Germany], enVista [Bausch & Lomb, Rochester, NY], and AcrySof IQ [Alcon Laboratories, Inc., Fort Worth, TX]) and one multifocal toric IOL (AcrySof IQ ReSTOR; Alcon Laboratories, Inc.) were implanted. Preoperative and postoperative images were taken to calculate the misalignment due to the marking method. To evaluate rotation in the different follow-up visits, another photograph was taken 1 hour and 1, 7 and 30 days postoperatively. Refraction, uncorrected distance visual acuity (UDVA), and corrected distance visual acuity were measured 30 days postoperatively. RESULTS: Postoperative UDVA was 0.1 logMAR or better in 64.6% of eyes implanted with monofocal IOLs and 46.4% of eyes implanted with multifocal IOLs. The enVista toric IOL showed the best UDVA compared to the other monofocal IOLs, with 81% of eyes with 0.1 logMAR or better. The mean misalignment in the total group studied was 0.07° ± 0.60°; 69.6% of monofocal IOLs and 67.9% of multifocal IOLs showed less than 5° of rotation. A correlation was found between postoperative UDVA and rotation in the monofocal and multifocal IOLs implanted (r = 0.439 [P < .011] and = 0.787 [P = .001], respectively). CONCLUSIONS: At 1 month postoperatively, UDVA was slightly more affected by IOL rotation in multifocal than monofocal toric IOLs. The marking method was also effective. | |
dc.description.department | Depto. de Optometría y Visión | |
dc.description.faculty | Fac. de Óptica y Optometría | |
dc.description.refereed | TRUE | |
dc.description.status | pub | |
dc.eprint.id | https://eprints.ucm.es/id/eprint/31303 | |
dc.identifier.doi | 10.3928/1081597X-20150122-03 | |
dc.identifier.issn | 1081-597X | |
dc.identifier.officialurl | http://dx.doi.org/10.3928/1081597X-20150122-03 | |
dc.identifier.relatedurl | http://www.healio.com/ophthalmology/journals/jrs | |
dc.identifier.uri | https://hdl.handle.net/20.500.14352/24076 | |
dc.issue.number | 2 | |
dc.journal.title | Journal of Refractive Surgery | |
dc.language.iso | eng | |
dc.page.final | 97 | |
dc.page.initial | 90 | |
dc.publisher | Slack Incorporated | |
dc.rights.accessRights | open access | |
dc.subject.cdu | 617.7-089.243-089 | |
dc.subject.keyword | Lens Implantation | |
dc.subject.keyword | Astigmatism | |
dc.subject.keyword | Cataract Surgery | |
dc.subject.keyword | Clinical outcomes | |
dc.subject.keyword | Silicone | |
dc.subject.keyword | Eyes | |
dc.subject.keyword | Shrinkage | |
dc.subject.keyword | Phacoemulsification | |
dc.subject.keyword | Visual Acuity | |
dc.subject.keyword | Prosthesis Desing | |
dc.subject.keyword | Pseudophakia | |
dc.subject.ucm | Cirugía | |
dc.subject.ucm | Oftalmología | |
dc.subject.ucm | Lentes de contacto | |
dc.subject.ucm | Óptica oftálmica | |
dc.subject.unesco | 3213 Cirugía | |
dc.subject.unesco | 3201.09 Oftalmología | |
dc.subject.unesco | 3311.11 Instrumentos ópticos | |
dc.title | Evaluation of Rotation and Visual Outcomes After Implantation of Monofocal and Multifocal Toric Intraocular Lenses | |
dc.type | journal article | |
dc.volume.number | 31 | |
dcterms.references | 1. Bachernegg A, Ruckl T, Riha W, Grabner G, Dexl AK. Rotational stability and visual outcome after implantation of a new toric intraocular lens for the correction of corneal astigmatism during cataract surgery. J Cataract Refract Surg. 2013;39:1390-1398. 2. Wolffsohn JS, Buckhurst PJ. Objective analysis of toric intraocular lens rotation and centration. J Cataract Refract Surg. 2010;36:778-782. 3. Felipe A, Artigas JM, Diez-Ajenjo A, Garcia-Domene C, Alcocer P. Residual astigmatism produced by toric intraocular lens rotation. J Cataract Refract Surg. 2011;37:1895-1901. 4. Jain AK, Gupta R, Chakma P. A simple technique for corneal marking before toric intraocular lens implantation. J Cataract Refract Surg. 2013;39:966-967. 5. Roumeliotis GA, Hutnik CM. A simple, reproducible, and cost effective axis marking system for toric lens implantation. J Refract Surg. 2012;28:12-13. 6. Visser N, Berendschot TT, Bauer NJ, Jurich J, Kersting O, Nuijts RM. Accuracy of toric intraocular lens implantation in cataract and refractive surgery. J Cataract Refract Surg. 2011;37:1394-1402. 7. Cha D, Kang SY, Kim SH, Song JS, Kim HM. New axis-marking method for a toric intraocular lens: mapping method. J Refract Surg. 2011;27:375-379. 8. Graether JM. Simplified system of marking the cornea for a toric intraocular lens. J Cataract Refract Surg. 2009;35:1498-1500. 9. Viestenz A, Seitz B, Langenbucher A. Evaluating the eye’s rotational stability during standard photography: effect on determining the axial orientation of toric intraocular lenses. J Cataract Refract Surg. 2005;31:557-561. 10. Popp N, Hirnschall N, Maedel S, Findl O. Evaluation of 4 corneal astigmatic marking methods. J Cataract Refract Surg. 2012;38:2094-2099. 11. Chua WH, Yuen LH, Chua J, Teh G, Hill WE. Matched comparison of rotational stability of 1-piece acrylic and plate-haptic silicone toric intraocular lenses in Asian eyes. J Cataract Refract Surg. 2012;38:620-624. 12. Prinz A, Neumayer T, Buehl W, et al. Rotational stability and posterior capsule opacification of a plate-haptic and an open-loop-haptic intraocular lens. J Cataract Refract Surg. 2011;37:251-257. 13. Buckhurst PJ, Wolffsohn JS, Naroo SA, Davies LN. Rotational and centration stability of an aspheric intraocular lens with a simulated toric design. J Cataract Refract Surg. 2010;36:1523-1528. 14. Ferreira TB, Marques EF, Rodrigues A, Montes-Mico R. Visual and optical outcomes of a diffractive multifocal toric intraocular lens. J Cataract Refract Surg. 2013;39:1029-1035. 15. Alfonso JF, Knorz M, Fernandez-Vega L, et al. Clinical outcomes after bilateral implantation of an apodized +3.0 D toric diffractive multifocal intraocular lens. J Cataract Refract Surg. 2014;40:51-59. 16. Bellucci R, Bauer NJ, Daya SM, et al. Visual acuity and refraction with a diffractive multifocal toric intraocular lens. J Cataract Refract Surg. 2013;39:1507-1518. 17. Williams JR. The Declaration of Helsinki and public health. Bull World Health Organ. 2008;86:650-652. 18. Pereira FA, Milverton EJ, Coroneo MT. Miyake-Apple study of the rotational stability of the AcrySof Toric intraocular lens after experimental eye trauma. Eye (Lond). 2010;24:376-378. 19. Chang DF. Early rotational stability of the longer Staar toric intraocular lens: fifty consecutive cases. J Cataract Refract Surg. 2003;29:935-940. 20. Ohmi S. Decentration associated with asymmetric capsular shrinkage and intraocular lens size. J Cataract Refract Surg. 1993;19:640-643. 21. Ohmi S, Tanaka S, Saika S, Ohnishi Y. Silicone and poly(methyl methacrylate) lens decentration associated with asymmetric capsule shrinkage in rabbits. J Cataract Refract Surg. 1999;25:1147-1150. 22. Khng C, Osher RH. Evaluation of the relationship between corneal diameter and lens diameter. J Cataract Refract Surg. 2008;34:475-479. 23. Patel CK, Ormonde S, Rosen PH, Bron AJ. Postoperative intraocular lens rotation: a randomized comparison of plate and loop haptic implants. Ophthalmology. 1999;106:2190-2195. 24. Stephenson M. Early users share the enVista experience. Review of Ophthalmology. 2013;January(available: http://www.reviewofophthalmology.com/content/t/intraocular_lenses/c/38573/) 25. Visser N, Bauer NJ, Nuijts RM. Toric intraocular lenses: historical overview, patient selection, IOL calculation, surgical techniques, clinical outcomes, and complications. J Cataract Refract Surg. 2013;39:624-637. 26. Mendicute J, Irigoyen C, Aramberri J, Ondarra A, Montés-Micó R. Foldable toric intraocular lens for astigmatism correction in cataract patients. J Cataract Refract Surg. 2008;34:601-607. 27. Dardzhikova A, Shah CR, Gimbel HV. Early experience with the AcrySof toric IOL for the correction of astigmatism in cataract surgery. Can J Ophthalmol. 2009;44:269-273. 28. Tsinopoulos IT, Tsaousis KT, Tsakpinis D, Ziakas NG, Dimitrakos SA. Acrylic toric intraocular lens implantation: a single center experience concerning clinical outcomes and postoperative rotation. Clin Ophthalmol. 2010;4:137-142. | |
dspace.entity.type | Publication | |
relation.isAuthorOfPublication | 367ce491-287f-4b2a-ba6a-870a393c5609 | |
relation.isAuthorOfPublication | 1c171089-8e25-448f-bcce-28d030f8f43a | |
relation.isAuthorOfPublication.latestForDiscovery | 367ce491-287f-4b2a-ba6a-870a393c5609 |
Download
Original bundle
1 - 1 of 1