Accuracy and Repeatability of Spatiotemporal Gait Parameters Measured with an Inertial Measurement Unit

dc.contributor.authorPosada Ordax, Jorge
dc.contributor.authorCosín Matamoros, Julia María
dc.contributor.authorLosa Iglesias, Marta Elena
dc.contributor.authorBecerro De Bengoa Vallejo, Ricardo
dc.contributor.authorEsteban Gonzalo, Laura
dc.contributor.authorMartín Villa, Carlos
dc.contributor.authorCalvo Lobo, César
dc.contributor.authorRodríguez Sanz, David
dc.date.accessioned2023-06-17T08:24:00Z
dc.date.available2023-06-17T08:24:00Z
dc.date.issued2021-04-21
dc.description.abstractIn recent years, interest in finding alternatives for the evaluation of mobility has increased. Inertial measurement units (IMUs) stand out for their portability, size, and low price. The objective of this study was to examine the accuracy and repeatability of a commercially available IMU under controlled conditions in healthy subjects. A total of 36 subjects, including 17 males and 19 females were analyzed with a Wiva Science IMU in a corridor test while walking for 10 m and in a threadmill at 1.6 km/h, 2.4 km/h, 3.2 km/h, 4 km/h, and 4.8 km/h for one minute. We found no difference when we compared the variables at 4 km/h and 4.8 km/h. However, we found greater differences and errors at 1.6 km/h, 2.4 km/h and 3.2 km/h, and the latter one (1.6 km/h) generated more error. The main conclusion is that the Wiva Science IMU is reliable at high speeds but loses reliability at low speeds.
dc.description.departmentDepto. de Enfermería
dc.description.facultyFac. de Enfermería, Fisioterapia y Podología
dc.description.refereedTRUE
dc.description.statuspub
dc.eprint.idhttps://eprints.ucm.es/id/eprint/71310
dc.identifier.doi10.3390/jcm10091804
dc.identifier.issn2077-0383
dc.identifier.officialurlhttps://doi.org/10.3390/jcm10091804
dc.identifier.relatedurlhttps://www.mdpi.com/2077-0383/10/9/1804/htm
dc.identifier.urihttps://hdl.handle.net/20.500.14352/6968
dc.issue.number9
dc.journal.titleJournal of Clinical Medicine
dc.language.isoeng
dc.page.initial1804
dc.publisherMPDI
dc.rightsAtribución 3.0 España
dc.rights.accessRightsopen access
dc.rights.urihttps://creativecommons.org/licenses/by/3.0/es/
dc.subject.keywordaccuracy
dc.subject.keywordrepeatability
dc.subject.keywordinertial
dc.subject.ucmMedicina
dc.subject.unesco32 Ciencias Médicas
dc.titleAccuracy and Repeatability of Spatiotemporal Gait Parameters Measured with an Inertial Measurement Unit
dc.typejournal article
dc.volume.number10
dspace.entity.typePublication
relation.isAuthorOfPublicationbbd621a9-7baf-481a-83c7-124287192d07
relation.isAuthorOfPublication6fc582ad-caa5-4262-bbb4-f58f9a31ee4f
relation.isAuthorOfPublication9420a0f4-5654-4ad9-a920-e9521d454023
relation.isAuthorOfPublication3d3e5c8a-15a9-4991-97ee-ea886940b1a5
relation.isAuthorOfPublication.latestForDiscovery6fc582ad-caa5-4262-bbb4-f58f9a31ee4f

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
jcm-10-01804.pdf
Size:
419.18 KB
Format:
Adobe Portable Document Format

Collections