Influence of arch location and scanning pattern on the scanning accuracy, scanning time, and number of photograms of complete-arch intraoral digital implant scans

dc.contributor.authorGómez Polo, Miguel Ángel
dc.contributor.authorCascos, Rocío
dc.contributor.authorOrtega, Rocío
dc.contributor.authorBarmak, Abdul Bassir
dc.contributor.authorKois, John C.
dc.contributor.authorRevilla-León, Marta
dc.date.accessioned2024-12-16T12:21:02Z
dc.date.available2024-12-16T12:21:02Z
dc.date.issued2023-03-29
dc.description.abstractObjectives: To measure the influence of arch location and scanning pattern on the accuracy, scanning time, and number of photograms of complete-arch implant scans acquired using an intraoral scanner (IOS). Materials and methods: A maxillary (maxillary group) and mandibular (mandibular group) model with 6 implant abutments on each cast was digitized using a desktop scanner (control scans). Six subgroups were created based on the scanning pattern used to acquire the scans using an IOS (Trios 4): occluso-buccal-lingual (OBL subgroup), occluso-linguo-buccal (OLB subgroup), bucco-linguo-occlusal (BLO subgroup), linguo-buccal-occlusal (LBO subgroup), zigzag (ZZ subgroup), and circumferential (C subgroup). The control scans were used as a reference to measure the discrepancy with the experimental scans calculating the root mean square error. Two-way ANOVA and the pairwise comparison Tukey tests were used to analyze the data (α = .05). Results: Significant discrepancies in trueness (p < .001), precision (p < .001), scanning time (p < .001), and number of photograms (p < .001) were found. The maxillary group obtained poorer trueness and precision values, higher scanning times, and a larger number of photograms than the mandibular group. The C subgroup obtained the best trueness and precision values, but was not significantly different from the OLB, BLO, and LBO subgroups. The ZZ subgroup obtained the worst trueness and precision values (p < .05). The C subgroup obtained the lowest scanning time and number of photograms (p < .05). Conclusions: Arch location and scanning pattern influenced scanning accuracy, scanning time, and number of photograms of complete-arch implant scans.
dc.description.departmentDepto. de Odontología Conservadora y Prótesis
dc.description.facultyFac. de Odontología
dc.description.refereedTRUE
dc.description.statuspub
dc.identifier.citationGómez-Polo M, Cascos R, Ortega R, Barmak AB, Kois JC, Revilla-León M. Influence of arch location and scanning pattern on the scanning accuracy, scanning time, and number of photograms of complete-arch intraoral digital implant scans. Clin Oral Implants Res. 2023 Jun;34(6):591-601. doi: 10.1111/clr.14069. Epub 2023 Apr 13. PMID: 37052054.
dc.identifier.doi10.1111/clr.14069
dc.identifier.essn1600-0501
dc.identifier.issn0905-7161
dc.identifier.officialurlhttps://onlinelibrary.wiley.com/journal/16000501
dc.identifier.pmid37052054
dc.identifier.urihttps://hdl.handle.net/20.500.14352/112665
dc.issue.number6
dc.journal.titleClinical Oral Implants Research
dc.language.isoeng
dc.page.final601
dc.page.initial591
dc.publisherWiley
dc.rights.accessRightsrestricted access
dc.subject.cdu616.314-089.843
dc.subject.keywordDental implants
dc.subject.keywordIntraoral scanners
dc.subject.keywordScanning accuracy
dc.subject.keywordScanning pattern
dc.subject.keywordScanning strategy
dc.subject.ucmImplantes dentales
dc.subject.ucmOdontología (Odontología)
dc.subject.unesco3299 Otras Especialidades Médicas
dc.subject.unesco3213.13 Ortodoncia-Estomatología
dc.titleInfluence of arch location and scanning pattern on the scanning accuracy, scanning time, and number of photograms of complete-arch intraoral digital implant scans
dc.typejournal article
dc.type.hasVersionAM
dc.volume.number34
dspace.entity.typePublication
relation.isAuthorOfPublication9700c70b-34f2-4d6a-8e12-92587c9186fb
relation.isAuthorOfPublication.latestForDiscovery9700c70b-34f2-4d6a-8e12-92587c9186fb

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
nfluence of arch location 2023 .pdf
Size:
6.58 MB
Format:
Adobe Portable Document Format

Collections