Validation of the use of a smart band in recording spatiotemporal gait parameters in the 6-minute walk test
| dc.contributor.author | Ortiz Gutiérrez, Rosa | |
| dc.contributor.author | López Marcos, José Javier | |
| dc.contributor.author | Maté Muñoz, José Luis | |
| dc.contributor.author | Moreta De Esteban, Paloma | |
| dc.contributor.author | Martín Casas, Patricia | |
| dc.date.accessioned | 2025-10-08T18:13:48Z | |
| dc.date.available | 2025-10-08T18:13:48Z | |
| dc.date.issued | 2025-04-21 | |
| dc.description | This article belongs to the Special Issue Sensors and Artificial Intelligence for Analyzing Human Behavior in Sports and Physical Activity | |
| dc.description.abstract | Wearable monitoring devices, such as smart bands, have emerged as accessible and non-invasive tools for assessing physiological and spatiotemporal gait parameters in various clinical tests. This study aimed to validate the use of the Xiaomi Mi Band 6 for recording gait parameters during the six-minute walk test (6MWT). Seventy participants without gait impairments were recruited, and the measurements obtained with the smart band were compared to reference methods (evaluator, pedometer, and pulse oximeter). The physiological parameter results showed that the smart band demonstrated good accuracy in heart rate monitoring but lower agreement in oxygen saturation measurements. Gait parameters indicated excellent agreement in step count (ICC > 0.9) and step frequency (ICC > 0.75), whereas step length and distance estimations showed higher variability. These findings suggest that the Xiaomi Mi Band 6 is a viable alternative for measuring specific gait parameters, though with limitations in certain aspects of accuracy. | |
| dc.description.department | Depto. de Fisioterapia | |
| dc.description.faculty | Fac. de Enfermería, Fisioterapia y Podología | |
| dc.description.refereed | TRUE | |
| dc.description.status | pub | |
| dc.identifier.citation | Ortiz-Gutiérrez, R.M.; López-Marcos, J.J.; Maté-Muñoz, J.L.; Moreta-de-Esteban, P.; Martín-Casas, P. Validation of the Use of a Smart Band in Recording Spatiotemporal Gait Parameters in the 6-Minute Walk Test. Sensors 2025, 25, 2621. https://doi.org/10.3390/s25082621 | |
| dc.identifier.doi | 10.3390/s25082621 | |
| dc.identifier.officialurl | https://doi.org/10.3390/s25082621 | |
| dc.identifier.relatedurl | https://www.mdpi.com/1424-8220/25/8/2621 | |
| dc.identifier.uri | https://hdl.handle.net/20.500.14352/124687 | |
| dc.issue.number | 8 | |
| dc.journal.title | Sensors | |
| dc.language.iso | eng | |
| dc.page.final | 15 | |
| dc.page.initial | 1 | |
| dc.publisher | Multidisciplinary Digital Publishing Institute (MDPI) | |
| dc.rights | Attribution 4.0 International | en |
| dc.rights.accessRights | open access | |
| dc.rights.uri | http://creativecommons.org/licenses/by/4.0/ | |
| dc.subject.cdu | 615.8 | |
| dc.subject.keyword | gait analysis | |
| dc.subject.keyword | wearable electronic devices | |
| dc.subject.keyword | digital health | |
| dc.subject.keyword | validation study | |
| dc.subject.ucm | Ciencias Biomédicas | |
| dc.subject.unesco | 3299 Otras Especialidades Médicas | |
| dc.title | Validation of the use of a smart band in recording spatiotemporal gait parameters in the 6-minute walk test | |
| dc.type | journal article | |
| dc.type.hasVersion | VoR | |
| dc.volume.number | 25 | |
| dspace.entity.type | Publication | |
| relation.isAuthorOfPublication | 254c9b60-c456-45d8-86a1-f16e2e375b12 | |
| relation.isAuthorOfPublication | 66b405c9-4ec9-4d29-96d3-7d62a046cbdc | |
| relation.isAuthorOfPublication | 940310d8-32ac-4af2-b6ee-a7c9eb72a6c0 | |
| relation.isAuthorOfPublication | 56e03ac7-ee70-40c5-88ed-daa491e647d9 | |
| relation.isAuthorOfPublication | affec10d-497c-4aae-8366-5f86b383b8c5 | |
| relation.isAuthorOfPublication.latestForDiscovery | 254c9b60-c456-45d8-86a1-f16e2e375b12 |
Download
Original bundle
1 - 1 of 1
Loading...
- Name:
- Validation_of_the_Use_Smart.pdf
- Size:
- 1.23 MB
- Format:
- Adobe Portable Document Format