ASTEROID stereotest v1.0: Lower stereo thresholds usingsmaller, denser and faster dots

dc.contributor.authorRead, Jenny
dc.contributor.authorWong, Zhen Yi
dc.contributor.authorYek, Xinye
dc.contributor.authorWong, Ying Xin
dc.contributor.authorBachtoula González, Omar
dc.contributor.authorLlamas Cornejo, Ichasus
dc.contributor.authorSerrano Pedraza, Ignacio
dc.date.accessioned2025-01-24T13:41:01Z
dc.date.available2025-01-24T13:41:01Z
dc.date.issued2020-09-29
dc.description.abstractPurpose: In 2019, we described ASTEROID, a new stereotest run on a 3D tabletcomputer which involves a four-alternative disparity detection task on a dynamicrandom-dot stereogram. Stereo thresholds measured with ASTEROID were wellcorrelated with, but systematically higher than (by a factor of around 1.5), thresh-olds measured with previous laboratory stereotests or the Randot Preschool clini-cal stereotest. We speculated that this might be due to the relatively large, sparsedots used in ASTEROID v0.9. Here, we introduce and test the stereo thresholdsand test-repeatability of the new ASTEROID v1.0, which uses precomputedimages to allow stereograms made up of much smaller, denser dots.Methods: Stereo thresholds and test/retest repeatability were tested and comparedbetween the old and new versions of ASTEROID (n = 75) and the Randot Circles(n = 31) stereotest, in healthy young adults.Results: Thresholds on ASTEROID v1.0 are lower (better) than on ASTEROIDv0.9 by a factor of 1.4, and do not differ significantly from thresholds on the Ran-dot Circles. Thresholds were roughly log-normally distributed with a mean of1.54 log 10 arcsec (35 arcsec) on ASTEROID v1.0 compared to 1.70 log 10 arcsec(50 arcsec) on ASTEROID v0.9. The standard deviation between observers wasthe same for both versions, 0.32 log 10 arcsec, corresponding to a factor of 2 aboveand below the mean. There was no difference between the versions in their test/retest repeatability, with 95% coefficient of repeatability = 0.46 log10 arcsec (afactor of 2.9 or 1.5 octaves) and a Pearson correlation of 0.8 (comparable to otherclinical stereotests).Conclusion: The poorer stereo thresholds previously reported with ASTEROIDv0.9 appear to have been due to the relatively large, coarse dots and low densityused, rather than to some other aspect of the technology. Employing the smalldots and high density used in ASTEROID v1.0, thresholds and test/retest repeata-bility are similar to other clinical stereotests.
dc.description.departmentDepto. de Psicología Experimental, Procesos Cognitivos y Logopedia
dc.description.facultyFac. de Psicología
dc.description.refereedTRUE
dc.description.sponsorshipMinisterio de Ciencia, Innovación y Universidades (España)
dc.description.statuspub
dc.identifier.citationRead, J. C. A., Wong, Z. Y., Yek, X., Wong, Y. X., Bachtoula, O., Llamas‐Cornejo, I., & Serrano‐Pedraza, I. (2020). ASTEROID stereotest v1.0: Lower stereo thresholds using smaller, denser and faster dots. Ophthalmic and Physiological Optics, 40(6), 815-827. https://doi.org/10.1111/opo.12737
dc.identifier.doi10.1111/opo.12737
dc.identifier.essn1475-1313
dc.identifier.issn0275-5408
dc.identifier.officialurlhttps://doi.org/10.1111/opo.12737
dc.identifier.relatedurlhttps://onlinelibrary.wiley.com/doi/10.1111/opo.12737
dc.identifier.urihttps://hdl.handle.net/20.500.14352/116051
dc.issue.number6
dc.journal.titleOphthalmic and Physiological Optics
dc.language.isoeng
dc.page.final827
dc.page.initial815
dc.publisherWiley on behalf of College of Optometrists
dc.relation.projectIDinfo:eu-repo/grantAgreement/AEI/Plan Estatal de Investigación Científica y Técnica y de Innovación 2017-2020/PGC2018-093406-B-I00/ES/LA INTERACCION BINOCULAR EN LA PERCEPCION DEL MOVIMIENTO: IMPLICACIONES EN LA ACTIVACION DE LOS MECANISMOS VISUALES INHIBITORIOS/
dc.rightsAttribution 4.0 Internationalen
dc.rights.accessRightsopen access
dc.rights.urihttp://creativecommons.org/licenses/by/4.0/
dc.subject.keywordBinocular vision
dc.subject.keywordPsychophysics
dc.subject.keywordStereoacuity
dc.subject.keywordStereopsis
dc.subject.keywordStereoscopic vision
dc.subject.keywordStereotest
dc.subject.keywordVision tests
dc.subject.ucmPercepción
dc.subject.unesco6106 Psicología Experimental
dc.titleASTEROID stereotest v1.0: Lower stereo thresholds usingsmaller, denser and faster dots
dc.typejournal article
dc.type.hasVersionVoR
dc.volume.number40
dspace.entity.typePublication
relation.isAuthorOfPublication6657e4c8-7806-4035-af4b-4df58d59ec8f
relation.isAuthorOfPublication09b6c2c0-a8bf-4cb6-b66b-580fa33d4bc5
relation.isAuthorOfPublication0fc94368-bbc4-426b-915d-34d2e98197db
relation.isAuthorOfPublication.latestForDiscovery0fc94368-bbc4-426b-915d-34d2e98197db

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
ReadEA2020.pdf
Size:
3.13 MB
Format:
Adobe Portable Document Format

Collections