Aviso: para depositar documentos, por favor, inicia sesión e identifícate con tu cuenta de correo institucional de la UCM con el botón MI CUENTA UCM. No emplees la opción AUTENTICACIÓN CON CONTRASEÑA
 

Agreement between the biometric measurements used to calculate the size of the implantable collamer lenses measured with four different technologies

dc.contributor.authorCalvo Sanz, Jorge Antonio
dc.contributor.authorPoyales Galán, Francisco
dc.contributor.authorZhou, Ying
dc.contributor.authorArias Puente, Alfonso
dc.contributor.authorGarzón Jiménez, Nuria
dc.date.accessioned2023-06-22T10:43:10Z
dc.date.available2023-06-22T10:43:10Z
dc.date.issued2022-04-28
dc.descriptionReceived: 26-Aug-2021; Revision: 07-Dec-2021; Accepted: 21-Jan-2022; Published: 28-Apr-2022
dc.description.abstractPurpose: To evaluate the agreement between the biometric measurements used to calculate the size of the implantable collamer lenses (ICL) with different technologies: swept-source optical coherence tomography, spectral domain optical coherence tomography, and Scheimpflug tomography. Methods: This retrospective observational study included subjects undergoing refractive surgery with posterior chamber phakic IOL implantation to correct their myopia. The anterior chamber depth (ACD) and the horizontal white to white (WTW) or the angle to angle (ATA) distance were measured with the following four devices: the IOLMaster 700 biometer (Carl Zeiss Meditec, Jena, Germany), based on swept-source optical coherence tomography; the Cirrus and Visante optical coherence tomographs (Carl Zeiss Meditec) based on low-coherence interferometry; and the Pentacam rotating Scheimpflug camera (Oculus, Wetzlar, Germany). Results: In the horizontal corneal diameter measurements, there were statistically significant differences between Pentacam–IOLMaster 700 pair (P < 0.001) and Pentacam–Visante pair (P < 0.001). WTW from CIRRUS showed the lowest correlation when paired with Pentacam and IOLMaster 700 (R2 = 0.452 and 0.385 for Visante and R2 = 0.494 and 0.426 for Cirrus). Regarding the linear correlation of the ACD measurements, all pairs of devices were statistically significant and all of them showed a very good correlation index. Conclusion: There is a good agreement between the different devices under evaluation for ACD measurements. As for WTW, the values measured with the different devices showed large discrepancies with low correlation levels, especially when comparing the tomographs with the other devices under evaluation.
dc.description.departmentDepto. de Optometría y Visión
dc.description.facultyFac. de Óptica y Optometría
dc.description.refereedTRUE
dc.description.statuspub
dc.eprint.idhttps://eprints.ucm.es/id/eprint/72447
dc.identifier.doi10.4103/ijo.IJO_2217_21
dc.identifier.issn0301-4738
dc.identifier.officialurlhttps://doi.org/10.4103/ijo.ijo_2217_21
dc.identifier.relatedurlhttps://journals.lww.com/ijo/Fulltext/2022/05000/Agreement_between_the_biometric_measurements_used.29.aspx
dc.identifier.urihttps://hdl.handle.net/20.500.14352/71495
dc.issue.number5
dc.journal.titleIndian Journal of Ophthalmology
dc.language.isoeng
dc.page.final1592
dc.page.initial1586
dc.publisherWolters Kluwer Medknow Publications
dc.rightsAtribución-NoComercial-CompartirIgual 3.0 España
dc.rights.accessRightsopen access
dc.rights.urihttps://creativecommons.org/licenses/by-nc-sa/3.0/es/
dc.subject.cdu681.73
dc.subject.cdu617.75-07
dc.subject.keywordAnterior Chamber Depth
dc.subject.keywordICL
dc.subject.keywordphakic lens
dc.subject.keywordwhite to white
dc.subject.ucmÓptica y optometría
dc.subject.ucmMateriales ópticos
dc.subject.ucmTécnicas de la imagen
dc.subject.unesco2209 Óptica
dc.subject.unesco2209.14 Propiedad ópticas de los sólidos
dc.titleAgreement between the biometric measurements used to calculate the size of the implantable collamer lenses measured with four different technologies
dc.typejournal article
dc.volume.number70
dspace.entity.typePublication
relation.isAuthorOfPublication367ce491-287f-4b2a-ba6a-870a393c5609
relation.isAuthorOfPublication.latestForDiscovery367ce491-287f-4b2a-ba6a-870a393c5609

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Agreement_between_the_biometric_measurements_used.29.pdf
Size:
1.26 MB
Format:
Adobe Portable Document Format

Collections