Aviso: para depositar documentos, por favor, inicia sesión e identifícate con tu cuenta de correo institucional de la UCM con el botón MI CUENTA UCM. No emplees la opción AUTENTICACIÓN CON CONTRASEÑA
 

Visual acuity of pseudophakic patients predicted from in-vitro measurements of intraocular lenses with different design

dc.contributor.authorVega, Fidel
dc.contributor.authorSagrario Millán, Maria
dc.contributor.authorGarzón Jiménez, Nuria
dc.contributor.authorAltemir, Irene
dc.contributor.authorPoyales, Francisco
dc.contributor.authorLarrosa, José Manuel
dc.date.accessioned2023-11-17T20:20:36Z
dc.date.available2023-11-17T20:20:36Z
dc.date.issued2018
dc.description.abstractThe optical quality of a set of IOLs (modeling set: one monofocal and two bifocals) was assessed through focus by the area under the modulation transfer function (MTFa) metric and related to the visual acuity (VA) defocus curves of pseudophakic patients implanted with said IOLs. A non-linear relationship between the MTFa and clinical VA was obtained with an asymptotic limit found to be the best VA achievable by the patients. Two mathematical fitting functions between clinical VA and MTFa were derived with high correlation coefficients (R2≥0.85). They were applied to the MTFa obtained from a different set of IOLs with advanced designs (trial set: one extended range of vision -ERV-, one trifocal ERV and one trifocal apodized) to predict VA versus defocus of patients implanted with these IOLs. Differences between the calculated VA and the clinical VA for both fitting models were within the standard deviation of the clinical measurements in the range of -3.00 D to 0.00 D defocus, thus proving the suitability of the MTFa metric to predict clinical VA performance of new IOL designs.
dc.description.departmentDepto. de Optometría y Visión
dc.description.facultyFac. de Óptica y Optometría
dc.description.refereedTRUE
dc.description.sponsorshipMinisterio de Economía y Competitividad (España)
dc.description.sponsorshipEuropean Commission
dc.description.statuspub
dc.identifier.citationVega F, Millán MS, Garzón N, Altemir I, Poyales F, Larrosa JM. Visual acuity of pseudophakic patients predicted from in-vitro measurements of intraocular lenses with different design. Biomed Opt Express. 2018 Sep 17;9(10):4893-4906.
dc.identifier.doi10.1364/boe.9.004893
dc.identifier.issn2156-7085
dc.identifier.issn2156-7085
dc.identifier.officialurlhttps://doi.org/10.1364/BOE.9.004893
dc.identifier.urihttps://hdl.handle.net/20.500.14352/88809
dc.issue.number10
dc.journal.titleBiomed Opt Express
dc.language.isoeng
dc.page.final4906
dc.page.initial4893
dc.publisherOptica Publishing Group
dc.relation.projectIDDPI2016-76019-R
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internationalen
dc.rights.accessRightsopen access
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/
dc.subject.cdu617.75
dc.subject.cdu617.7-089.243
dc.subject.cdu681.73
dc.subject.keywordDiffractive lenses; Ophthalmic optics and devices; Visual acuity; Visual optics, modeling; Refractive surgery.
dc.subject.ucmÓptica y optometría
dc.subject.ucmCirugía
dc.subject.ucmMateriales ópticos
dc.subject.unesco3213.09 Cirugía Ocular
dc.subject.unesco3311.09 Lentes
dc.subject.unesco2209.15 Optometría
dc.titleVisual acuity of pseudophakic patients predicted from in-vitro measurements of intraocular lenses with different design
dc.typejournal article
dc.type.hasVersionAM
dc.volume.number9
dspace.entity.typePublication
relation.isAuthorOfPublication367ce491-287f-4b2a-ba6a-870a393c5609
relation.isAuthorOfPublication.latestForDiscovery367ce491-287f-4b2a-ba6a-870a393c5609

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Visual acuity of pseudophakic patients predicted from in-vitro measurements of intraocular lenses with different design.pdf
Size:
4.44 MB
Format:
Adobe Portable Document Format

Collections