Aviso: para depositar documentos, por favor, inicia sesión e identifícate con tu cuenta de correo institucional de la UCM con el botón MI CUENTA UCM. No emplees la opción AUTENTICACIÓN CON CONTRASEÑA
 

Determination of the oxygen permeability (Dk) of contact lenses with a fiber-optic luminescent sensor system

dc.contributor.authorPérez Ortíz, Noelia
dc.contributor.authorNavarro Villoslada, Fernando
dc.contributor.authorOrellana Moraleda, Guillermo
dc.contributor.authorMoreno Jiménez, Florencio
dc.date.accessioned2023-06-20T12:55:56Z
dc.date.available2023-06-20T12:55:56Z
dc.date.issued2007-03-30
dc.descriptionReceived 27 December 2006, Accepted 20 March 2007, Available online 30 March 2007.
dc.description.abstractAn optical sensing system, alternative to current electrochemical techniques, has been developed for determining the oxygen permeability (Dk) of rigid gas-permeable contact lenses. The novel sensor is based on kinetic measurements of the oxygen partial pressure inside a chamber sealed by the sample contact lens, where a thin luminescent O2-sensitive film is placed. The standard equation for calculating the Dk value of the lens material has been adapted to the optical technique and a fitting procedure of the sensor response has been developed to achieve the maximum accuracy and precision. Unlike the electrochemical techniques, the optical sensor is unaffected by the contact lens thickness and the lens edge and boundary effects. Therefore, the latter does not require lengthy measuring procedures for extracting the actual Dk values. The optical sensor system has been applied to oxygen permeability determinations of commercial contact lenses of nominal Dk values equal to 17 and 36 ANSI units of 10−11 cm3 cm/s cm2 mmHg.
dc.description.departmentDepto. de Química Orgánica
dc.description.departmentSección Deptal. de Química Orgánica (Óptica y Optometría)
dc.description.facultyFac. de Ciencias Químicas
dc.description.facultyFac. de Óptica y Optometría
dc.description.refereedTRUE
dc.description.sponsorshipComunidad de Madrid/ FEDER
dc.description.statuspub
dc.eprint.idhttps://eprints.ucm.es/id/eprint/71165
dc.identifier.doidoi.org/10.1016/j.snb.2007.03.027
dc.identifier.issn0925-4005
dc.identifier.officialurlhttps://doi.org/10.1016/j.snb.2007.03.027
dc.identifier.urihttps://hdl.handle.net/20.500.14352/52833
dc.issue.number2
dc.journal.titleSensors and Actuators B - Chemical
dc.language.isoeng
dc.page.final399
dc.page.initial394
dc.publisherElsevier Science S A
dc.relation.projectIDCM-PRICYT(S-0505/AMB/0374)
dc.rights.accessRightsrestricted access
dc.subject.cdu617.7‑089.243
dc.subject.cdu539.1.073.5
dc.subject.cdu681.73
dc.subject.keywordOxygen permeability
dc.subject.keywordContact lenses
dc.subject.keywordDk
dc.subject.keywordOptical sensor
dc.subject.keywordLuminescence
dc.subject.ucmQuímica orgánica (Química)
dc.subject.ucmLentes de contacto
dc.subject.ucmMateriales ópticos
dc.subject.unesco2306 Química Orgánica
dc.subject.unesco3311.11 Instrumentos ópticos
dc.subject.unesco2209.14 Propiedad ópticas de los sólidos
dc.titleDetermination of the oxygen permeability (Dk) of contact lenses with a fiber-optic luminescent sensor system
dc.typejournal article
dc.volume.number126
dspace.entity.typePublication
relation.isAuthorOfPublicationb5c09973-d256-4f73-97ef-b8f218437cc3
relation.isAuthorOfPublicationdefd6c32-fdda-4eae-8e60-5942fcbed64b
relation.isAuthorOfPublication09111c59-3829-4b92-a9d2-ee099b545e9e
relation.isAuthorOfPublication.latestForDiscoveryb5c09973-d256-4f73-97ef-b8f218437cc3

Download

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
1-s2.0-S0925400507002079-main.pdf
Size:
305.11 KB
Format:
Adobe Portable Document Format

Collections