Dipole density of low-frequency and spike magnetic activity: A reliable procedure in presurgical evaluation of temporal lobe epilepsy
Loading...
Official URL
Full text at PDC
Publication date
2004
Advisors (or tutors)
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
American Clinical Neurophysiology Society
Citation
Fernández Lucas, A. A., De Sola, R. G., Amo, C. et al. «Dipole Density of Low-Frequency and Spike Magnetic Activity: A Reliable Procedure in Presurgical Evaluation of Temporal Lobe Epilepsy»: Journal of Clinical Neurophysiology, vol. 21, n.o 4, julio de 2004, pp. 254-66. DOI.org (Crossref), https://doi.org/10.1097/00004691-200407000-00004.
Abstract
Conventional visual analysis and dipole density analysis of magnetoencephalographic data for both spike and low-frequency magnetic activity were compared for presurgical evaluation in temporal lobe epilepsy (TLE) in a sample of 26 drug-resistant operated TLE patients. A series of logistic regression analyses were performed. Dipole density sensitivity was superior to visual localization analysis. Three separated logistic models were calculated for interictal spikes, low-frequency magnetic activity, and the combination of both measures. A combined interictal spike/low-frequency magnetic activity model predicted correctly the operated temporal lobe in all patients. Clear-cut criteria for the probability model are proposed that are valid for 92.3% of cases in the sample. The quantitative approach proposed by this study is an evidence-based model for presurgical evaluation of temporal lobe epilepsy, which improves previous magnetoencephalographic investigations and establishes working clinical criteria for patient evaluation in TLE.