Para depositar en Docta Complutense, identifícate con tu correo @ucm.es en el SSO institucional: Haz clic en el desplegable de INICIO DE SESIÓN situado en la parte superior derecha de la pantalla. Introduce tu correo electrónico y tu contraseña de la UCM y haz clic en el botón MI CUENTA UCM, no autenticación con contraseña.
 

Agreement between central corneal thickness measured using pentacam, ultrasound pachymetry, specular microscopy and optic biometer lenstar LS 900 and the influence of intraocular pressure

Citation

Borrego Sanz, L., Saenz Frances, F., Bermudez Vallecilla, M. C. et al. «Agreement between Central Corneal Thickness Measured Using Pentacam, Ultrasound Pachymetry, Specular Microscopy and Optic Biometer Lenstar LS 900 and the Influence of Intraocular Pressure». Ophthalmologica, vol. 231, n.o 4, marzo de 2014, pp. 226-35. Silverchair, https://doi.org/10.1159/000356724.

Abstract

Purpose: To compare central corneal thickness (CCT) values obtained by Lenstar (LE), Pentacam (PC), specular microscopy (SM) and ultrasound pachymetry (UP) in healthy corneas and study their influence on intraocular pressure (IOP) readings determined by Goldmann applanation tonometry (GAT). Methods: CCT was measured in 76 healthy subjects by LE, PC, SM and UP. We established Lin's concordance correlation coefficient (ρ-C) between different techniques. The influence of CCT on GAT was established through univariate linear regression models, IOP being the dependent variable. Results: The highest ρ-C was found between LE and SM at 0.94 (95% CI: 0.91-0.96) and between LE and UP at 0.95 (95% CI: 0.94-0.97). IOP readings showed less variability when CCT was determined using LE (7.7%, B = 0.16; 95% CI: 0.004-0.28). Conclusions: Although CCT values obtained with UP, PC, SM and LE show good correlation, these methods are not completely interchangeable. The amount of IOP variation differs when CCT is determined using LE or SM.

Research Projects

Organizational Units

Journal Issue

Description

Keywords

Collections