Zero reference signal for displacement measuring systems by use of speckle

Loading...
Thumbnail Image

Full text at PDC

Publication date

2003

Advisors (or tutors)

Editors

Journal Title

Journal ISSN

Volume Title

Publisher

The Optical Society Of America
Citations
Google Scholar

Citation

Abstract

An optical method for generating a reference signal for an incremental displacement measurement system is proposed. We achieved this zero reference signal by comparing two speckle patterns arriving from two symmetric diffusers, which are used as natural random codes that are identical only when the reading head is located equidistantly between the diffusers. The comparison of the speckles is obtained either by interference, as in a Michelson interferometer, or by intensity correlations.

Research Projects

Organizational Units

Journal Issue

Description

© 2003 Optical Society of America. We thank M. Quintanilla for useful comments and discussions. This research has been supported by projects DPI 2001-1238 and DPI 2001-1369 of the Spanish Ministry for Science and Technology.