Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry

Loading...
Thumbnail Image
Full text at PDC
Publication Date
2015
Authors
Cacho, Isabel
Sánchez Naves, Juan
Pintor Just, Jesús
Advisors (or tutors)
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
Hindawi Publishing Corporation
Citations
Google Scholar
Research Projects
Organizational Units
Journal Issue
Abstract
Purpose. To compare the intraocular pressure (IOP) before and after Laser In Situ Keratomileusis (LASIK), measured by Diaton, Perkins, and noncontact air pulse tonometers. Methods. Fifty-seven patients with a mean age of 34.88 were scheduled for myopia LASIK treatment. Spherical equivalent refraction (SER), corneal curvature (K), and central corneal thickness (CCT) and superior corneal thickness (SCT) were obtained before and after LASIK surgery. IOP values before and after surgery were measured using Diaton, Perkins, and noncontact air pulse tonometers. Results. The IOP values before and after LASIK surgery using Perkins tonometer and air tonometers were statistically significant (). However, no significant differences were found () for IOP values measured with Diaton tonometer. CCT decreases significantly after surgery () but no statistical differences were found in SCT (). Correlations between pre- and postsurgery were found for all tonometers used, with and for the air pulse tonometer, and for Perkins, and and for Diaton. Conclusion. Transpalpebral tonometry may be useful for measuring postsurgery IOP after myopic LASIK ablation because this technique is not influenced by the treatment.
Description
En Acceso abierto en la web del editor. Received 19 March 2015; Revised 28 May 2015; Accepted 28 May 2015
Keywords
Citation
- M. Kohlhaas, E. Spoerl, A. G. Boehm, and K. Pollack, “A correction formula for the real intraocular pressure after LASIK for the correction of myopic astigmatism,” Journal of Refractive Surgery, vol. 22, no. 3, pp. 263–267, 2006. - J. Sánchez-Navés, L. Furfaro, O. Piro, and S. Balle, “Impact and permanence of LASIK-induced structural changes in the cornea on pneumotonometric measurements: contributions of flap cutting and stromal ablation,” Journal of Glaucoma, vol. 17, no. 8, pp. 611–618, 2008. View at Publisher · View at Google Scholar· - L. M. Agudelo, C. A. Molina, and D. L. Alvarez, “Changes in intraocular pressure after laser in situ keratomileusis for myopia, hyperopia, and astigmatism,” Journal of Refractive Surgery, vol. 18, no. 4, pp. 472–474, 2002. - H. A. Quigley, “Glaucoma,” The Lancet, vol. 377, no. 9774, pp. 1367–1377, 2011. - M. Qiu, S. Y. Wang, K. Singh, and S. C. Lin, “Association between myopia and glaucoma in the united states population,” Investigative Ophthalmology and Visual Science, vol. 54, no. 1, pp. 830–835, 2013. - T. Y. Wong, B. E. K. Klein, R. Klein, M. Knudtson, and K. E. Lee, “Refractive errors, intraocular pressure, and glaucoma in a white population,” Ophthalmology, vol. 110, no. 1, pp. 211–217, 2003. - S. A. Perera, T. Y. Wong, W.-T. Tay, P. J. Foster, S.-M. Saw, and T. Aung, “Refractive error, axial dimensions, and primary open-angle glaucoma: the Singapore Malay Eye Study,” Archives of Ophthalmology, vol. 128, no. 7, pp. 900–905, 2010. - S.-M. Saw, G. Gazzard, E. C. Shin-Yen, and W.-H. Chua, “Myopia and associated pathological complications,” Ophthalmic and Physiological Optics, vol. 25, no. 5, pp. 381–391, 2005. - M. L. Conway, M. Wevill, A. Benavente-Perez, and S. L. Hosking, “Ocular blood-flow hemodynamics before and after application of a laser in situ keratomileusis ring,” Journal of Cataract and Refractive Surgery, vol. 36, no. 2, pp. 268–272, 2010. - D. Ortiz, D. Piñero, M. H. Shabayek, F. Arnalich-Montiel, and J. L. Alió, “Corneal biomechanical properties in normal, post-laser in situ keratomileusis, and keratoconic eyes,” Journal of Cataract and Refractive Surgery, vol. 33, no. 8, pp. 1371–1375, 2007. - D. A. Luce, “Determining in vivo biomechanical properties of the cornea with an ocular response analyzer,” Journal of Cataract and Refractive Surgery, vol. 31, no. 1, pp. 156–162, 2005. - A. Bhan, A. C. Browning, S. Shah, R. Hamilton, D. Dave, and H. S. Dua, “Effect of corneal thickness on intraocular pressure measurements with the pneumotonometer, Goldmann applanation tonometer, and Tono-Pen,” Investigative Ophthalmology and Visual Science, vol. 43, no. 5, pp. 1389–1392, 2002. - J. Liu and C. J. Roberts, “Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis,” Journal of Cataract and Refractive Surgery, vol. 31, no. 1, pp. 146–155, 2005. - A. T. Broman, N. G. Congdon, K. Bandeen-Roche, and H. A. Quigley, “Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure,” Journal of Glaucoma, vol. 16, no. 7, pp. 581–588, 2007. - P. Ceruti, R. Morbio, M. Marraffa, and G. Marchini, “Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyes,” Eye, vol. 23, no. 2, pp. 262–269, 2009. - E. Schneider and F. Grehn, “Intraocular pressure measurement—comparison of dynamic contour tonometry and goldmann applanation tonometry,” Journal of Glaucoma, vol. 15, no. 1, pp. 2–6, 2006. - M. A. El Danasoury, A. El Maghraby, and S. J. Coorpender, “Change in intraocular pressure in myopic eyes measured with contact and non-contact tonometers after laser in situ keratomileusis,” Journal of Refractive Surgery, vol. 17, no. 2, pp. 97–104, 2001. - M. I. Toker, A. Vural, H. Erdogan, A. Topalkara, and M. K. Arici, “Central corneal thickness and Diaton transpalpebral tonometry,” Graefe's Archive for Clinical and Experimental Ophthalmology, vol. 246, no. 6, pp. 881–889, 2008. - C. G. V. De Moraes, T. S. Prata, J. Liebmann, and R. Ritch, “Modalities of tonometry and their accuracy with respect to corneal thickness and irregularities,” Journal of Optometry, vol. 1, no. 2, pp. 43–49, 2008. - World Medical Association, “World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects,” The Journal of the American Medical Association, vol. 310, pp. 2191–3194, 2013. - A. S. H. Tsai and S. C. Loon, “Intraocular pressure assessment after laser in situ keratomileusis: a review,” Clinical and Experimental Ophthalmology, vol. 40, no. 3, pp. 295–304, 2012. - K. Yi, G. Bae, M. Kong, and E.-S. Chung, “Intraocular pressure measured with goldmann, noncontact, schiøtz, and dynamic contour tonometry after DSEK,” Cornea, vol. 32, no. 8, pp. 1089–1093, 2013. - S. Sudesh, M. J. Moseley, and J. R. Thompson, “Accuracy of Goldmann tonometry in clinical practice,” Acta Ophthalmologica, vol. 71, no. 2, pp. 185–188, 1993. - A. P. Nesterov, G. K. Piletskiĭ, and N. G. Piletskiĭ, “Transpalpebral tonometer for measuring intraocular pressure,” Vestnik Oftalmologii, vol. 119, no. 1, pp. 3–5, 2003. - A. Lösch, A. Scheuerle, V. Rupp, G. Auffarth, and M. Becker, “Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry,” Graefe's Archive for Clinical and Experimental Ophthalmology, vol. 243, no. 4, pp. 313–316, 2005. - D. Sandner, A. Böhm, S. Kostov, and L. Pillunat, “Measurement of the intraocular pressure with the ‘transpalpebral tonometer’ TGDc-01 in comparison with applanation tonometry,” Graefe's Archive for Clinical and Experimental Ophthalmology, vol. 243, no. 6, pp. 563–569, 2005. - J. A. Cook, A. P. Botello, A. Elders et al., “Systematic review of the agreement of tonometers with Goldmann applanation tonometry,” Ophthalmology, vol. 119, no. 8, pp. 1552–1557, 2012. - M. D. Doherty, Z. I. Carrim, and D. P. O'Neill, “Diaton tonometry: an assessment of validity and preference against Goldmann tonometry,” Clinical and Experimental Ophthalmology, vol. 40, no. 4, pp. e171–e175, 2012. - G. Shemesh, U. Soiberman, and S. Kurtz, “Intraocular pressure measurements with Goldmann applanation tonometry and dynamic contour tonometry in eyes after IntraLASIK or LASEK,” Clinical Ophthalmology, vol. 6, no. 1, pp. 1967–1969, 2012. - S. M. A. Shousha, M. A. Abo Steit, M. H. Hosny, W. A. Ewais, and A. M. Shalaby, “Comparison of different intraocular pressure measurement techniques in normal eyes, post surface and post lamellar refractive surgery,” Clinical Ophthalmology, vol. 7, no. 1, pp. 71–79, 2013. - M. Kohlhaas, E. Spörl, A. G. Böhm, K. Pollack, D. Sandner, and L. E. Pillunat, “Applanation tonometry in ‘normal’ patients and patients after LASIK,” Klinische Monatsblatter fur Augenheilkunde, vol. 222, no. 10, pp. 823–826, 2005. - T. Schlote and H. Landenberger, “Intraocular pressure difference in Goldmann applanation tonometry versus a transpalpebral tonometer TGDc-01 ‘PRA’ in glaucoma patients,” Klinische Monatsblatter fur Augenheilkunde, vol. 222, no. 2, pp. 123–131, 2005.
Collections