Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry
dc.contributor.author | Cacho, Isabel | |
dc.contributor.author | Sánchez Naves, Juan | |
dc.contributor.author | Batres Valderas, Laura | |
dc.contributor.author | Carracedo Rodríguez, Juan Gonzalo | |
dc.contributor.author | Pintor Just, Jesús Jerónimo | |
dc.date | Received 19 March 2015; Revised 28 May 2015; Accepted 28 May 2015 | |
dc.date.accessioned | 2023-06-19T15:03:31Z | |
dc.date.available | 2023-06-19T15:03:31Z | |
dc.date.issued | 2015 | |
dc.description | En Acceso abierto en la web del editor. | |
dc.description.abstract | Purpose. To compare the intraocular pressure (IOP) before and after Laser In Situ Keratomileusis (LASIK), measured by Diaton, Perkins, and noncontact air pulse tonometers. Methods. Fifty-seven patients with a mean age of 34.88 were scheduled for myopia LASIK treatment. Spherical equivalent refraction (SER), corneal curvature (K), and central corneal thickness (CCT) and superior corneal thickness (SCT) were obtained before and after LASIK surgery. IOP values before and after surgery were measured using Diaton, Perkins, and noncontact air pulse tonometers. Results. The IOP values before and after LASIK surgery using Perkins tonometer and air tonometers were statistically significant (). However, no significant differences were found () for IOP values measured with Diaton tonometer. CCT decreases significantly after surgery () but no statistical differences were found in SCT (). Correlations between pre- and postsurgery were found for all tonometers used, with and for the air pulse tonometer, and for Perkins, and and for Diaton. Conclusion. Transpalpebral tonometry may be useful for measuring postsurgery IOP after myopic LASIK ablation because this technique is not influenced by the treatment. | en |
dc.description.department | Depto. de Optometría y Visión | |
dc.description.department | Unidad Docente de Bioquímica y Biología Molecular | |
dc.description.faculty | Fac. de Óptica y Optometría | |
dc.description.refereed | TRUE | |
dc.description.sponsorship | Ministerio de Ciencia, Innovación y Universidades (España) | |
dc.description.sponsorship | Redes Temáticas de Investigación Cooperativa en Salud (España) | |
dc.description.status | pub | |
dc.eprint.id | https://eprints.ucm.es/id/eprint/39209 | |
dc.identifier.citation | Cacho, I., Sánchez Navas, J., Batres Valderas, L. et al. «Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry». Journal of Ophthalmology, vol. 2015, 2015, pp. 1-6. DOI.org (Crossref), https://doi.org/10.1155/2015/683895. | |
dc.identifier.doi | 10.1155/2015/683895 | |
dc.identifier.issn | 2090-004X | |
dc.identifier.officialurl | https://dx.doi.org/10.1155/2015/683895 | |
dc.identifier.relatedurl | https://www.hindawi.com/journals/joph/2015/683895/ | |
dc.identifier.uri | https://hdl.handle.net/20.500.14352/35235 | |
dc.issue.number | 683895 | |
dc.journal.title | Journal of Ophthalmology | |
dc.language.iso | eng | |
dc.page.total | 6 | |
dc.publisher | Hindawi Publishing Corporation | |
dc.relation.projectID | SAF2013-44416-R | |
dc.relation.projectID | RD12/0034/0003 | |
dc.rights | Atribución 3.0 España | |
dc.rights.accessRights | open access | |
dc.rights.uri | https://creativecommons.org/licenses/by/3.0/es/ | |
dc.subject.cdu | 617.713-089 | |
dc.subject.cdu | 617.7-089.243 | |
dc.subject.keyword | Intraocular lenses | |
dc.subject.keyword | Laser surgery | |
dc.subject.keyword | Keratomileusis | |
dc.subject.keyword | Tonometry | |
dc.subject.ucm | Cirugía | |
dc.subject.ucm | Oftalmología | |
dc.subject.ucm | Lentes de contacto | |
dc.subject.unesco | 3213 Cirugía | |
dc.subject.unesco | 3201.09 Oftalmología | |
dc.subject.unesco | 3311.11 Instrumentos ópticos | |
dc.title | Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry | en |
dc.type | journal article | |
dc.volume.number | 2015 | |
dspace.entity.type | Publication | |
relation.isAuthorOfPublication | 43c348f6-e5ef-4c0d-944c-bf6098098eab | |
relation.isAuthorOfPublication | 97433353-d31b-4cf5-b2c0-47d2e6703fe5 | |
relation.isAuthorOfPublication | e8366c14-6aee-427c-8601-f6bf1e360010 | |
relation.isAuthorOfPublication.latestForDiscovery | 43c348f6-e5ef-4c0d-944c-bf6098098eab |
Download
Original bundle