A sensitive and rapid method for the determination of the clinically relevant biomarker receptor tyrosine kinase AXL in serum involving amperometric disposable immunosensors is reported. The target protein was sandwiched between a specific capture antibody covalently immobilized on screen-printed carbon electrodes modified with electropolymerized poly(pyrrolepropionic acid) and a biotinylated detector antibody labeled with a streptavidin-horseradish peroxidase conjugate. The amperometric responses were measured at −0.20 V vs the Ag pseudo-reference electrode of the SPCE upon the addition of H2O2 in the presence of hydroquinone (HQ) as mediator. This integrated immunosensing platform showed a low limit of detection (337 pg mL−1), a good selectivity against other non-target serum proteins, and provided results statistically in agreement with those obtained by using a commercial ELISA kit. These attractive features together with the simplicity and easy automation and miniaturization of the required instrumentation make the developed methodology a promising alternative in the development of devices for on-site clinical analysis.

Research Projects

Organizational Units

Journal Issue

Description

Keywords

Collections