Influence of Instrumental Factors in the Measurement of Power Profiles of Intraocular Lenses with a Commercial Deflectometer

Loading...
Thumbnail Image

Full text at PDC

Publication date

2023

Advisors (or tutors)

Editors

Journal Title

Journal ISSN

Volume Title

Publisher

MDPI
Citations
Google Scholar

Citation

Gómez Pedrero, J. A., Albarrán Diego, C., García Montero, M. et al. «Influence of Instrumental Factors in the Measurement of Power Profiles of Intraocular Lenses with a Commercial Deflectometer». Applied Sciences, vol. 13, n.o 17, agosto de 2023, p. 9882. DOI.org (Crossref), https://doi.org/10.3390/app13179882.

Abstract

Deflectometry is an optical technique for determining properties such as power distribution, wavefront, etc., and measurement of the optical properties of an intraocular lens can provide relevant information for clinicians. The aim of the current study was to establish a protocol for measuring lens power maps and profiles of various optical designs of intraocular lenses with a deflectometer based on the phase-shifting Schlieren principle (NIMO TR1504, Lambda-X, Nivelles, Belgium). The results are discussed with respect to accuracy and repeatability, the influence of the use of filters, and whether to consider the intraocular lens as a thin or thick lens.