High performance of the automated ADVIA Centaur Systems SARS-CoV-2 Antigen Assay in nasopharyngeal samples with high viral load
Loading...
Official URL
Full text at PDC
Publication date
2022
Advisors (or tutors)
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
Springer Nature
Citation
Ríos, E., Medrano, S., Alvarez, M., Valderrama, M. J., Vallejo, L., Delgado-Iribarren, A., & Culebras, E. (2023). High performance of the automated ADVIA Centaur Systems SARS-CoV-2 Antigen Assay in nasopharyngeal samples with high viral load. International Microbiology, 26(3), 471-474. https://doi.org/10.1007/S10123-022-00311-3
Abstract
ADVIA Centaur SARS-CoV-2 Antigen (COV2Ag) Assay (Siemens Healthineers) was evaluated for SARS-CoV-2 detection. A total of 141 nasopharyngeal samples were analyzed by this technique and results were compared with those obtained by quantitative reverse-transcription polymerase chain reaction (RT-PCR). The overall sensitivity and specificity of the test were 68.70% and 70%, respectively. Regarding cycle threshold (Ct) values, the COV2Ag test showed a sensitivity of 93.75% and 100% for nasopharyngeal samples with Ct < 25 and < 20, respectively. ADVIA Centaur COV2Ag Assay is a useful, automated, and rapid technique for early SARS-CoV-2 diagnosis and isolation of the infected individuals, avoiding its transmission.
Description
This study was supported by the Community of Madrid and the European Union, through the European Regional Development Fund (ERDF), supported as part of the Union’s response to the COVID-19 pandemic (PROYECTO REACT-EU INMUNOVACTER-CM).